6-5
IM 04P01B01-01E
Setup Operations for Convenient Functions (Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
6.4 Setting Recording Zones for Each Channel
(Zone Recording)
This section explains the details of setting the recording zone for each measurement
channel.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key or
SHIFT
+
key to select
Aux
and then press the key.
3.
Press the
key or
SHIFT
+
key to select
Zone
and then press the key.
4.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the
Setting complete
screen appears, do either of the following:
Press the key to set other channels.
If you are done, press the
ESC
key.
6.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Set the channel range.
Set the leftmost value of the
recording zone.
Set the rightmost value of the
recording zone.
The new setting takes effect.
First channel
Last channel
ESC/?
Set=Aux
Aux=Zone
CH=01-01
Left= 0
Right=100
01-01 Channel
Setting complete
Explanation
Zone
The recording zone can be set in the following range.
Leftmost position: 0 mm to 95 mm
Rightmost position: 5 mm to 100 mm
Set a larger value for the rightmost position than the leftmost position.
Set at least 5 mm for the zone width.
Note
If the zone is set to a value less than 40 mm, scale printout of periodic printout cannot be
executed.
Содержание mR10000 436101
Страница 1: ...User s Manual Model 436101 436102 436103 436104 436106 µR10000 Recorder IM 04P01B01 01E 8th Edition ...
Страница 2: ......
Страница 14: ...Blank ...
Страница 18: ...Blank ...
Страница 58: ...Blank ...
Страница 72: ...Blank ...
Страница 92: ...Blank ...
Страница 220: ...Blank ...
Страница 268: ...Blank ...