IM 12D8K4-E-E

4th edition

Instruction 
Manual

Model ISC40PR
Retractable fitting for
Inductive Conductivity sensor 

Содержание ISC40PR

Страница 1: ...IM 12D8K4 E E 4th edition Instruction Manual Model ISC40PR Retractable fitting for Inductive Conductivity sensor ...

Страница 2: ... 2 Installation site 6 3 3 Safety precautions 6 3 4 Installation method 6 3 5 Assembly of accessories 6 3 6 Dimensions 7 3 7 Dimensions of the options 8 3 8 Installation examples 9 3 9 Mounting the sensor into the fitting 10 3 10 Probe insertion 11 3 11 Adjusting the insertion depth 13 4 MAINTENANCE 14 4 1 General 14 4 2 Taking out the sensor 14 4 3 Replacing the sensor 14 4 4 Drain port connectio...

Страница 3: ...ption of the process G Adjustable insertion depth for optimal measuring position in the process G Replacement of the sensor G Improved sealing and damping by additional o rings G Simplified installation by optional ball valves with flanged or tapered connections G Available in 0 5 and 1 metre probe length in two different materials 1 3 Warranty Yokogawa Europe B V warrants that the goods delivered...

Страница 4: ... can be adjusted to a maximum of either 0 5 or 1 metre The actual probe insertion length depends on the use of optional adapters and or ball valves The dimensional drawing indicates the maximum insertion length for both specified probe lengths Pressure temperature ratings Static conditions 2 MPa at 20 C 500 kPa at 130 C Operating conditions maintenance max 500 kPa max 130 C Flange ratings DIN flan...

Страница 5: ...steel AISI 316 T Titanium Grade 2 or 3 Probe 05 0 5 m length 10 1 0 m A Adapter R2 G2 Stainless steel B Adapter 2 NPT G2 Stainless steel C Flange adapter Stainless steel D Flange adapter for drain Stainless steel E Adapter R2 G2 Titanium Grade 2 or 3 F Adapter 2 NPT G2 Titanium Grade 2 or 3 G Flange adapter Titanium Grade 2 or 3 H Flange adapter for drain Titanium Grade 2 or 3 M Material certifica...

Страница 6: ...tly 3 4 Installation method It is important to have the point of measurement in a location that is truly representing the process composition Check whether the specifications of the sensor fulfil the maximum occurring process conditions The fitting has several optional connection possibilities Check that you received the correct size and type Install the fitting in a convenient location for mainte...

Страница 7: ... 345 or 845 13 58 or 33 26 ISO 228 1 G2 140 5 51 min 150 5 9 max 130 5 11 max 160 or 660 6 29 or 25 98 min 150 5 9 NW50 150mm or 5 9 2 ANSI 180mm or 7 08 max NW50 120mm or 620 4 72 or 24 4 max 130 5 11 2 90mm or 590 3 54 or 23 22 Note Bolt nuts and gaskets are not included ...

Страница 8: ... G2 54 2 12 UNIVERSAL FLANGE DIN Flange DN50 PN16 ANSI Flange 2 150 lbs JIS Flange 2 150 lbs Options A B E F ISO 228 1 G2 ISO 7 1 R2 or 2 11 5 NPT Options D H ISO 228 1 G2 77 3 1 8 27 NPT UNIVERSAL FLANGE DIN Flange DN50 PN16 ANSI Flange 2 150 lbs JIS Flange 2 150 lbs ...

Страница 9: ...3 8 Installation examples IM 12D8K4 E E 9 ...

Страница 10: ...o the part in the metal tube 6 Lead the sensor cable through the tube of the fitting from the side where the knurled knob has been removed 5 4 12D8K4 07 SCREWING INTO TUBE 6 Hold the sensor still and turn the metal tube onto the sensor Don t rotate the cell but rotate the tube of the fitting because the cable can be disconnected from the cell when rotating it 7 Lead the loose part of the pigtail o...

Страница 11: ...rning the T bar key can only be done when the valve is closed FIXING SCREW 1 Pushing the probe into the process needs a force to overcome the pressure of the system and the friction of the dampening rings in the fitting The locking mechanism can be tightened until the probe is firmly fixed at the measuring position The insertion stop can be fixed at the actual insertion position See 3 10 on adjust...

Страница 12: ...IM 12D8K4 E E 12 4 3 5 ...

Страница 13: ...ed to your preference The insertion stop can be set for at the actual insertion position of the probe The Allan key supplied with the fitting is used to fix the insertion stop at the desired position The maximum insertion depth is related to the probe length mentioned in the model code of the fitting ...

Страница 14: ...valve 4 Turn the T bar counter clockwise 5 Take out the probe If the option drain port is available the process pressure can be relieved before removing the sensor Remarks Stand clear when releasing the fixing screw Due to the process pressure the probe can be pressed out The T bar key can only be operated when the ball valve is closed Make sure it is closed completely The friction of the o rings ...

Страница 15: ...520DD 14 Fixing screws 2 pcs 15 Chamber 16 1x Kalrez 70 shore 47 2x3 5 K1500EC 17 Adapter optional 18 Ball valve optional 19 Sensor 20 2 pcs Kalrez 70 shore 15 6x1 78 K1500ED 20 10 pcs Viton 70 shore 15 6x1 78 K1500BE Description Part No O ring Kalrez 29 74x3 53 K1500AH O ring set EPDM ISC40 PR K1500EA Sensor holder SS ISC40PR K1500EN Sensor holder Ti ISC40PR K1500EP Outer tube 0 5m SS ISC40PR K15...

Страница 16: ...253 7000 Fax 1 770 251 2088 E mail info yca com www yokogawa com us YOKOGAWA ELECTRIC ASIA Pte Ltd 5 Bedok South Road Singapore 469270 Singapore Tel 65 241 9933 Fax 65 241 2606 E mail webinfo yas com sg www yokogawa com sg YOKOGAWA HEADQUARTERS 9 32 Nakacho 2 chome Musashinoshi Tokyo 180 Japan Tel 81 422 52 5535 Fax 81 422 55 1202 E mail webinfo mls yokogawa co jp www yokogawa com Yokogawa has an ...