IM CA71-E
8-1
Communication Function
8
8. Communication Function - CA71 only -
You can configure the calibrator from a personal computer just as you
do with the calibrator's panel keys (except for turning on/off the power,
configuring the function selector switch, and setting the communication
function). You can also verify the setpoint, measured value and status
of the calibrator.
NOTE
• With the optional communication cable (P/N: 91017), you can use the RS232
serial port (9-pin D-sub) of a personal computer or any other equipment.
• In talk-only or printer mode, you can output the source setpoint and mea-
sured value at preset intervals.
8.1
Cables Connection and Interface Specifi-
cations
◆
Connecting Communication Cable
Remove the I/O port cover at the back of the calibrator and connect the
communication cable (P/N: 91017) to the I/O port.
◆
Setting RS232 Parameters
Baud rate:
9600 baud
Parity:
None
Stop bits:
2
Data length:
8 bits
Flow control:
None (Xon/Xoff control for printing only)
Delimiter:
Fixed to CrLf