7-4
IM CA71-E
7.2
Auto Step Function
The auto step function automatically changes the variable
n of the n/m
output in a step-by-step manner, as shown in the following figure, when
the divided output function (
n/m) is selected.
Setpoint
Sourced-value reading
Actual output
OFF
Stepping time setpoint
SOURCE ON
key operation
Press
(ON)
1
2
2
1
0
2
3
3
n
m
Auto Step Operation (when the default of
n is set to 1)
Step 1: Press the
key to turn off the calibrator.
Step 2: Place switch 1 (Sweep switch) in the ON (right-side) position.
Step 3: By setting the position of switch 2 (Speed switch), set the time
of one step.
OFF (left-side): 2.5 sec/step; ON (right-side): 5 sec/step
Step 4: Press the
key to turn on the calibrator.
Step 5: Using the function selector switch, select the function (volt-
age, current, resistance, thermocouple, or RTD) for which you
want to source a signal. In the case of pulse source, the auto
step function is disabled.
Step 6: Set the output value.
Step 7: Press the
key. The
indication on the LCD changes
to
and the
n/m symbol turns on.
7.2 Auto Step Function