IM CA71-E
12-1
Specifications
12
12. Specifications
Parameter
DC voltage
DC current
mA SINK
Resistance
RTD
TC *
4
Frequency,
pulse
Reference
100 mV
1 V
10 V
30 V
20 mA
4–20 mA
20 mA
400
Ω
Pt100 *
2
JPt100
K
E
J
T
N
L
U
R
S
B
500 Hz
1000 Hz
10 kHz
Pulse
cycle *
5
Range
-10.00–110.00 mV
0–1.1000 V
0–11.000 V
0–30.00 V
0–24.000 mA
4/8/12/16/20 mA
0.1–24.000 mA
0–400.00
Ω
-200.0–850.0
°
C
-200.0–500.0
°
C
-200.0–1372.0
°
C
-200.0–1000.0
°
C
-200.0–1200.0
°
C
-200.0–400.0
°
C
-200.0–1300.0
°
C
-200.0–900.0
°
C
-200.0–400.0
°
C
0–1768
°
C
600–1800
°
C
1.0–500.0 Hz
90–1100 Hz
0.9 kHz–11.0 kHz
1–99,999 cycles
Resolution
10
µ
V
0.1 mV
1 mV
10 mV
1
µ
A
4 mA
1
µ
A
0.01
Ω
0.1
°
C
0.1
°
C
1
°
C
0.1 Hz
1 Hz
0.1 kHz
1 cycle
Accuracy (23
±
5
°
C per year)
±
(0.02% + 15
µ
V)
±
(0.02% + 0.1 mV)
±
(0.02% + 1 mV)
±
(0.02% + 10 mV)
±
(0.025% + 3
µ
A)
±
(0.05% + 3
µ
A)
±
(0.025% + 0.1
Ω
)
±
(0.025% + 0.3
°
C)
±
(0.02% + 0.5
°
C)
(-100
°
C or greater)
±
(0.02% + 1
°
C)
(-100
°
C or less)
±
(0.02% + 0.5
°
C)
(0
°
C or greater)
±
(0.02% + 1
°
C)
(0
°
C or less)
±
(0.02% + 2.5
°
C)
(100
°
C or less)
±
(0.02% + 1.5
°
C)
(100
°
C or greater)
±
(0.02% + 2
°
C)
(1000
°
C or less)
±
(0.02% + 1.5
°
C)
(1000
°
C or greater)
±
0.2 Hz
±
1 Hz
±
0.1 kHz
–
Remarks
Maximum output: 5 mA
Maximum output: 10 mA
Maximum output: 10 mA *
1
Maximum load: 12 V
External power supply: 5–28 V
Excitation current: 0.5–5 mA *
3
If 0.1 mA, add 0.25
Ω
or 0.6
°
C. Subject
device input capacitance: 0.1
µ
F or less
TC source accuracy does not include RJ
sensor accuracy.
RJ sensor specs
Measurement range: -10–50
°
C
Accuracy (when combined with main unit)
18–28
°
C:
±
0.5
°
C
Other than the above:
±
1
°
C
Output voltage: +0.1–+15 V (zero base waveform)
Amplitude accuracy:
±
(5% + 0.1 V)
Maximum load current: 10 mA
Contact output
(with 0.0 V amplitude setting, FET switch
ON/OFF)
Maximum open/close voltage/current: +28 V/50
mA
Temperature coefficient: Accuracy shown above
×
(1/5)/
°
C
*1: Output up to 24 V/22 mA is possible when using the AC adapter.
*2: As per JIS C 1604-1997 (ITS-90). IPTS-68 may be selected through internal settings (DIP switch).
*3: Excitation current: If less than 0.1 mA to 0.5 mA, then add [0.025/ls (mA)]
Ω
or [0.06/ls (mA)]
°
C.
*4: As per JIS C 1602-1995 (ITS-90) (L and U are DIN specs).
K, E, J, T, N, R, S, and B may be switched to IPTS-68 through internal settings (DIP switch) (L and U are not switched).
*5: Frequency (interval between one pulse and another) and amplitude during pulse cycle source may have the same range as during
frequency source.
(1) Signal sourcing unit range and accuracy (for both CA51 and CA71)
±
(setting percentage plus
µ
V, mV, mA,
Ω
or
°
C)
Содержание HANDY CAL CA51
Страница 1: ...User s Manual Model CA51 71 HANDY CAL Calibrators IM CA71 E 1st Edition Feb 2002 IM CA71 E...
Страница 11: ......
Страница 29: ......
Страница 51: ......
Страница 61: ......
Страница 69: ......
Страница 77: ......
Страница 89: ......
Страница 91: ......
Страница 101: ......
Страница 103: ......