11.7 Error Log Display
Procedure
Error Log Display
1.
Press
ERROR
to display the error log screen.
Error message or
confirmation message
Clearing the Error Log Display
2.
Press the
Clear
soft key to clear the error log.
Explanation
The ERROR key illuminates if an error such as a runtime error or communication
command error is stored in the error memory.
The error log displays error messages and confirmation messages. An error message
has an associated error number, and a confirmation message does not.
A confirmation message is displayed in the error log until the problem indicated by the
message is resolved. It is not cleared even if the Clear soft key is pressed.
For the meaning and corrective action of messages, see section 17.2, “Error Code
Descriptions and Corrective Actions.”
Note
• The error memory is also cleared if the error is read using a communication command
“
:SYSTem:ERRor?
” or if the GS820 is turned OFF. You can also clear the error memory by
sending the “
*CLS
” communication command.
• If the error memory becomes full, “
−
350 : Queue overflow
” is entered as the last error
indicating that any addition errors will be discarded. If the error memory is cleared, the
ERROR key turns OFF.
<<Corresponding Command Mnemonic>>
:SYSTem:ERRor?
*CLS
11-8
IM 765601-01E