![YOKOGAWA FLXA402T Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/yokogawa/flxa402t/flxa402t_user-manual_911266063.webp)
<5. Others (Converter)>
5-6
IM 12A01G01-03EN
1st Edition : Mar. 25, 2021-00
5.1.1
Load individual configuration File
Configuration files for FLXA402T converter or sensors can be loaded individually as current
value.
Go to Converter menu > Others > Save/Load > Load individ. config. file
Select one below to see the list of configuration files.
Figure 5.4
Load Individ. config. file
To load individual configuration file, tap one file in the list.
Figure 5.5
Load individ. config. file
If you tap one of the files in the list above, a dialog box appears. (left below) In “File name”, you
will see the file name you selected. (left below) You cannot modify the file names.
On a dialog for loading sensor configuration file (right below), open the drop-down list next to
“Target sensor” to select one sensor connection. The drop-down list shows blank if there is no
valid sensor connection to load. “Execute” key does not work either.
Figure 5.6 sample (for converter)
Sample (for pH)
Confirm the file you selected is displayed in the box of File name (left above). Tap “Execute.” The
file starts being loaded and configuration is modified.
When Converter configuration file is loaded successfully, FLXA402T returns to Home screen.
When sensor configuration file is successfully loaded, FLXA402T returns to the screen of Load
individ. config. file, where you can select another file loading.
Содержание FLXA402T
Страница 13: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 57: ...Blank Page ...
Страница 75: ...Blank Page ...