4.8 Sweep Setting and Manipulation
97
IM FG410-01EN
4.8.5 Setting phase sweep
For the manipulation methods that are common with the sweep setting screen, refer to
pp. 83 and 86.
a) Phase sweep example
The following is an example of continuous, linear shuttle sweep.
Shuttle sweep
As the phase increases, the frequency rises by the value below, and as the phase decreases,
the frequency drops by the value below.
b) Selecting phase sweep
When [Mode] (oscillation mode) is set to
[Sweep] (sweep), set [Type] (sweep type)
on the 2nd page of the setting screen to
[Phase] (phase).
c) Waveforms for which phase sweep is not possible
Phase sweep is not possible for noise and DC.
d) Setting items required for phase sweep
Set the following items on the 2nd page of the setting screen. The phase settings on the 1st
page of the setting screen are disabled.
• Start phase [Start]
• Stop phase [Stop]
• Sweep time [Time]
This is the transition time from the start phase to the stop phase.
• Sweep mode [SwpMode]
Continuous, single-shot, and gated single-shot can be chosen from.
Sweep time
Sweep time
Sweep value
Output signal
Start value
Stop value
|
Stop phase (deg)
-
Start phase (deg)
|
1
Sweep time (s)
360
×
Select [Phase] in
[Type] and then
press the ENTER
key.
Содержание FG400
Страница 1: ...User s Manual FG400 Arbitrary Function Generator User s Manual Basic IM FG410 01EN 1st Edition ...
Страница 8: ...VI IM FG410 01EN ...
Страница 159: ...7 17 General Characteristics 147 IM FG410 01EN External dimensions FG410 Unit mm Rear view ...
Страница 160: ...148 7 SPECIFICATIONS IM FG410 01EN External dimensions FG420 Unit mm Rear view ...
Страница 161: ...RACK MOUNTING DIMENSIONS 149 IM FG410 01EN Inch rack mounting dimensions for 1 unit Unit mm RACK MOUNTING DIMENSIONS ...
Страница 163: ...RACK MOUNTING DIMENSIONS 151 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm ...
Страница 169: ...157 IM FG410 01EN ...
Страница 171: ......