< 3. Installation, Piping and Wiring >
3-2
IM 12F5A1-01E
3.2 Piping
3.2.1 For Use without a Sampling System
When sample has a pressure of 1 to 150 kPa and a flow rate of 0.1 to 2.5 l/min, such sample can be
directly introduced into the measuring tank. Connect a Ø6 X Ø4 mm polyethylene tube to the sample
water inlet.
Connect a Ø22 X Ø15 mm flexible mesh-reinforced tube to the drain outlet from the measuring tank.
Install this tube so that no water entrapment is formed. Entrapment of water in the drain tube may
cause overflow from the measuring tank.
F3.2e.EPS
Electrical connections
JIS A15 plastic water-
proof plug or equivalent
Drain outlet
(Ø22 x Ø15 mm hose coupling)
Measuring tank
Sample inlet
(Rc 1/4)
2B Pipe
Stanchion
Needle valve
Figure 3.2
Free Available Chlorine Analyzer Piping Connections