<2. Specifications>
2-18
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
2.5
ZA8F Flow Setting Unit and
ZR40H Automatic Calibration Unit
2.5.1
ZA8F Flow Setting Unit
This flow setting unit is applied to the reference gas and the calibration gas in a system
configuration (System 2). Used when instrument air is provided.
This unit consists of a flowmeter and flow control valves to control the flow rates of calibration gas
and reference gas.
Standard Specifications
Flowmeter Scale:
Calibration gas; 0.1 to 1.0 l/min. Reference gas; 0.1 to 1.0 l/min.
Construction:
Dust-proof and rainproof construction
Case Material:
SPCC (Cold rolled steel sheet)
Painting:
Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections:
Rc1/4 or 1/4FNPT
Reference Gas pressure: Clean air supply of sample gas pressure plus approx. 50 kPaG (or
sample gas pressure plus approx. 150 kPa G when a check valve is
used). Pressure at inlet of the Flow Setting Unit.(Maximum 300 kPaG)
Air Consumption:
Approx. 1.5 l/min
Weight:
Approx. 2.3 kg
Calibration gas (zero gas, span gas) Consumption: Approx. 0.7 l/min (at calibration time only)
NOTE
Use instrument air for span calibration gas, if no instrument air is available, contact YOKOGAWA.
l
Model and Codes
Model
Suffix code
Option code
Description
ZA8F
- - - - - - - -
- - - - - - - - -
Flow setting unit
Joint
-J
-A
- - - - - - - - -
- - - - - - - - -
Rc 1/4
With 1/4 NPT adapter
Style code
*C
- - - - - - - - -
Style C
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...