<8. Detailed Data Setting>
8-6
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
8.2.3
Output Hold Setting
To set the output hold, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Commissioning” in the Execution/Setup display. Next, select the “mA-output
setup” and then the “mA-outputs presets” display as shown in Figure 8.2.
Warm up: 4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Preset value: 4 . 0 mA
Cal.blow back: Hold
Preset value: 4 . 0 mA
Error: Preset
Preset value: 3 . 4 mA
Enter
mA-outputs presets
F8.1E.ai
Warm up: 4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Preset value: 4.0mA
Cal.blow back: Hold
Preset value: 4 . 0 mA
Error: Preset
Preset value: 3 . 4 mA
Enter
mA-outputs presets
Non-Hold
Hold
Preset
F8.2E.aiS
Figure 8.1
mA-outputs presets Display
Figure 8.2
Selecting Maintenance
(2) From this display (Fig. 8.2), select the desired display. Figure 8.3 shows an example
of selecting “Maintenance”. Select the desired output status.
(3) If a preset value is selected, set the corresponding output current. If you select a preset
value just below Maintenance on the screen, the numeric-data entry display will appear.
Enter the current value you want. To set 10 mA, type in 010 and press the [Enter] key to
complete the setting. The setting range is from 2.4 to 21.6 mA.
NOTE
“Error” of mA-outputs presets is not displayed when option code “/C2” or “/C3” (NAMUR NE 43
compliant) is specified.
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as shown in
Table 8.3.
Table 8.3
Output Hold Default Values
Status
Output hold setting
Preset value
During warm-up
4 mA
4 mA
Under maintenance
Holds output at value just before maintenance started.
4 mA
Under calibration or blow back Holds output at value just before starting calibration or
blow back.
4 mA
On Error occurrence
Holds output at a preset value.
3.4 mA
8.3
Alarms Setting
The analyzer enables the setting of four alarms — high-high, high, low, and low-low alarms
— depending upon the measurement conditions. The following section sets out the alarm
operations and setting procedures.
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...