
<7. Startup>
7-5
IM 11M12A01-02E
7.8.2
Humidity Analyzer - Analog Output Setting
Select any one of the analog output settings — Oxygen, Humidity, and Mixing from the mA-
output range display. If the /HS option is specified at the time of purchase, the equipment is a
humidity analyzer. For other than this setting, the analyzer is an oxygen analyzer. If a mixed
measurement is required, change the current output setting as given below. If the humidity
analyzer is specified in the above detector type setting, the analog output is set to “humidity”
when the data initialization is attempted.
(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select “mA-output setup”; the display shown in
Figure 7.9 then appears.
(3) Select “mA-output1” from the “mA-outputs” display. The “mA-output1 range” display shown
in Figure 7.10 then appears.
(4) Select the Parameter, then the Oxygen, Humidity, and Mixing appear in the mAoutput1
range display (Figure 7.11). Select “Mixing” and press the [Enter] key.
(5) The “Mixing” in Figure 7.12 appears.
(6) Perform the same setting in mA-outpout2 if necessary.
Both outputs are 4-20mA
mA-output1
mA-output2
Set presets
Enter
mA-outputs
F7-7E.ai
Parameter: Humidity
Set range
Min. humidity conc:
0 %H
2
O
Max. humidity conc:
25 %H
2
O
Output damping: 0 s
Output mode: Linear
Enter
mA-output1 range
F7-8E.ai
Figure 7.9
mA-outputs Setting Display
Figure 7.10
mA-output1 Range Display
Parameter: Humidity
Set range
Min. humidity conc:
0 %H2O
Max. humidity conc:
25 %H2O
Output damping: 0 s
Output mode: Linear
Enter
mA-output1 range
Oxygen
Humidity
Mixing
F7-9E.ai
Parameter: Mixing
Set range
Min. mixing ratio:
0. 000 kg/kg
Max. mixing ratio:
0. 200 kg/kg
Output damping: 0 s
Output mode: Linear
Enter
mA-output1 range
F7-10E.ai
Figure 7.11
mA-output1 Range Display
Figure 7.12
Mixing Ratio Display
7.8.3
Humidity Analyzer - Minimum Current (4 mA) and
Maximum Current (20 mA) Settings
To set the analog output range, follow these steps:
(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select the “mA-output setup.” Select “mAoutput1”
from the “mA-outputs” display. The “mA-output1 range” that displays “Parameter: Humidity,”
as shown in Figure 7.10 then appears.
16th Edition : Mar. 19, 2020-00
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...