The matrix can be cleared before entering new
values. Next new matrix values can be entered
as described above. The EXAxt can interpolate
the matrix. During this process it will check if
the matrix is completely ascending/descend-
ing. This is necessary as otherwise the lookup
function can give two results for one tempera-
ture. If an error is found, the EXAxt will specify
the location of the error as shown in the user
interface screen “user defined 1/2”.
The backspace key should be used for delet-
ing an individual matrix value.
An empty value is shown as
APPENDIX 2
Temperature compensation matrix.
1.
A minimum number of values is required to make interpolation possible.
The highlighted values marked as
are mandatory to enter.
Concent
Tref
T1
....
Tx
....
T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
SxTx
....
Sol10
C10
S10Tr
S10T1
S10T10
2.
Tref (reference temperature) is defined in the Temperature Compensation menu. If Tref is
between T1 and T10 then the value of Tref needs to be entered as Tx (T2....T9).
Concent
Tref
T1
....
Tx
.... T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
....
Sol10
C10
S10Tr
S10T1
S10T10
3.
For every SxTx that is entered the following values become mandatory to enter:
Cx, SxTr, SxT1, SxT10 and Tx.
Concent
Tref
T1
....
Tx
.... T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
Cx
SxTr
SxT1
SxTx
SxT10
....
Sol10
C10
S10Tr
S10T1
S10T10
Содержание EXAxt 450
Страница 1: ...IM 12D06D05 01E 4th Edition User s Manual Model ISC450G Style S2 Inductive Conductivity Converter ...
Страница 8: ...vi ...
Страница 23: ...15 IM 12D06D05 01E 3 INSTALLATION AND WIRING Figure 3 11 a Figure 3 11 b Figure 3 11 c ...
Страница 54: ...46 IM 12D06D05 01E ...
Страница 55: ...47 IM 12D06D05 01E 10 SPARE PARTS 10 SPARE PARTS See Customer Maintenance Parts List ...
Страница 66: ...58 IM 12D06D05 01E APPENDIX 6 Control drawing for FM approval ...