<6. PROCEDURES FOR CONVERTER OPERATION>
6-4
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
6.2.1 Functions in <MEAS.> Mode
Table 6.3 gives details on the functions in the <MEAS.> mode.
Table 6.3 Function in <MEAS.> Mode
F
Indication
Display range
Example of
indication
Initial (default)
value
Unit
Remarks
1 Turbidity
-10.00 to 2200
150.0
mg/L Degree/FTU is also
possible. *3
2 Lamp voltage
-0.100 to 6.000
3.600
V
3
Turbidity element
detection current
-99.99 to 5600 *1
1000
nA
4 Reference element
detection current
-0.2000 to 2.4000 *2
1.6000
V
5
Analog output in %
-10.0 to 110.0
50.0
%
6 Analog output in mA
2.40 to 21.60
12.00
mA
7
Zero-point error
-30.00 to 30.00
0.02
0.00
mg/L Degree/FTU is also
possible. *3
8 Slope
-50.0 to 400.0
150.0
100.0
%
9
|
F.
(Not to be selected)
*1, *2: These are indicated by reversing the sign (+ or -) for the voltage values obtained by placing the AG terminal of the detector to the
negative side (-).
*3:
"FTU" unit, a turbidity unit when formazin standard is used, is also available. Since the instrument uses kaolin standard,
reference sensitivity calibration using formazin standard is required for the FTU display.
Note 1: The values in the column for the initial value are those shown when the memory storing the data is initialized.
These are partially different from the values set upon shipment from the factory.