<3. INSTALLATION, PIPING, AND WIRING>
3-4
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
3.2.2 TB400G-
¨
-
¨
-A1, -A2, and -A3 (with sampling system)
(1) Sample-water piping
This is piping to supply sample to the detector. The pressure and flow rate for the Sample are 20
to 500 kPa and 2 to 10 L/min, respectively. Piping connections are designed for a rigid PVC pipe
of nominal size VP16 (ø22 mm). Do the piping using any flanges or unions conforming to the pipe
diameter.
(2) Tap-water piping
This is piping to supply cleaning water and zero water to the detector. Supply water having a
turbidity of 2 mg/L or less, such as tap water. The pressure of the water to be used is 100 to 500
kPa. Piping connections are designed for a rigid PVC pipe of nominal size VP16 (ø22 mm). Do
the piping using any fittings conforming to the pipe diameter similar to the piping for sample.
Note: Please consider to install a check valve so that cleaning water does not flow backward in the tap water side.
(3) Drain piping
This is piping for draining the sample and tap water supplied to the detector to a drain or the like.
The piping connection is designed for a rigid PVC pipe of nominal size VP40.
Connect such a pipe of nominal size VP40 (VU40) or more and do the piping so that debris does
not remain or get trapped or get trapped in the pipe.
Drain outlet
VP40 pipe
Tap water inlet
VP16 pipe
Sample inlet
VP16 pipe
F0304.ai
Figure 3.4 Piping Connections for Turbidimeter with Sampling System