<3. INSTALLATION, PIPING, AND WIRING>
3-1
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
3. INSTALLATION, PIPING, AND WIRING
3.1 Installation
3.1.1 Unpacking
The Surface Scattering Light Turbidity Meter is shipped only after being packed so as not to be
damaged during transportation. When the turbidimeter is delivered, unpack the meter carefully. If
the turbidimeter with a sampling system is delivered, unpack them near the location where they
are to be installed.
3.1.2 Installation Location
The Surface Scattering Light Turbidity Meter should be installed in a location where:
(1) rainwater cannot get inside, such as in a building or a cabinet,
(2) there is little vibration,
(3) corrosive gases are not present,
(4) there is low humidity,
(5) temperature changes are kept to a minimum and the temperature is maintained between -5
to 50 °C,
(6) there is enough clearance for maintenance and maintenance work can be easily done,
(7) drainage can be provided.