<6. Calibration Procesure>
6-2
IM 12J05D02-01E
l
Manual calibration
This is a calibration method whereby the sensor is not exposed to a calibration fluid, but
is calibrated by comparison with a reference method. This reference method is normally a
laboratory analysis of the water sample: either by iodometric titration according ISO 5813 or an
electrochemical method according ISO 5814.
The calibration is done in MAN.CAL mode in maintenance level.
NOTE
Calibration must be carried out after changing the manual pressure setting (see service code 51).
When the manual pressure is enabled, air and water calibrations are made based on the
specified pressure.
M.PR.ON YES/NO is displayed before the calibration to confirm whether the calibration will be
made based on the pressure.
6.1.2
Diagnostic functions performed during calibration
The calibration is a semi-automatic calibration, which means that the sensor output is used for
calculation of sensor parameters after the readings have stabilized. The criteria for stabilization
are set in service code 20. If stable readings are not achieved within one hour error E1 message
will appear on the display and the procedure is aborted. The calculated sensor parameters are
compared with preset tolerances and if these conditions are not met, Error E2 (Zero Error) or E3
(Slope Error) will be signalled.
NOTE
1. Error messages that appear initially indicate inadequate settings of the diagnostic variables
in service codes 01, 20 and/or 22.
2. Patience improves the accuracy of calibration. When the EXA prompts the user that the
preset criteria are met by showing CAL.END message it only means that the minimum
criteria are met. When the signal still changes after this initial acceptance, EXA will
automatically update the sensor characteristics with the new information.
6.2
Calibration procedure using air calibration
method
6.2.1 Preparation
Move the sensor to a maintenance site and wash off any dirt on the membrane. Lightly wipe off
any remaining water from the membrane with a soft tissue.
Expose the sensor to the air in an environment where there is no temperature change and no
wind. (A convenient way to accomplish this environment is to place the sensor in an empty
bucket)
NOTE
If sensors are used without temperature compensating element, the temperature must be
adjusted to ambient temperature in service code 13.
Before zero calibration, see 6.3.1 Preparation
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...