<4.  Maintenance>

4-1

IM 12J05D04-01E

4. Maintenance

4.1 

Routine Inspection

n

  Cleaning of Sensor (Sensor Cap)

A visual inspection of the sensor cap must be made whenever the sensor is calibrated. Any dirt 

on the sensor cap adversely affects your measurements.
If the dirt is spread over the sensor cap, dip the sensor cap into clean water and gently wipe 

the dirt off with a soft cloth or lint-free paper. If the dirt does not come off easily, apply a neutral 

detergent to the sensor cap so it can be easier to remove the dirt. Wash the sensor cap with 

clean water.
Before every calibration, check that the sensor cap is not loose. If it is loose, retighten it. When 

you retighten or repalce the sensor cap, use an attached silicon tube and so forth.

Sensor Cap

(*): When you attach or detach the sensor cap, use an attached silicon tube and so forth.

Silicon Tube (

*)

4.2 

Inspection If Failure Occurs

When the sensor cap ages and comes to the end of its service life, troubles such as failed span 

adjustment may occur. DO402G Dissolved Oxygen Converter or FLXA402 4-Wire Converter 

detects these failures.
If a sensor error code appears on the display, follow one of the procedures below.
1.  Clean the sensor (sensor cap), by following the direction of 4.1 above.
2.  Replace the sensor cap and the attached O-ring together with new ones.
3.  Replace DO70G sensor unit with a new one if resistance temperature detector (RTD) fails 

or the error still occurs even after the procedure 1 or 2 above is applied.

 

When DO402G is connected, troubles may be found by measuring the resistance value of 

RTD.

 

Another possibility is that the error may be caused by defects of RTD. 

 

A NTC 22k resistance temperature detector (RTD) is incorporated in the DO70G sensor. 

This RTD indicates a resistance value of about 22 kΩ at 25°C. Check the RTD by measuring 

the resistance value between the white wiring and the green wiring. If the resistance value 

differs greatly from the 22 kΩ, the sensor must be replaced.

For further information on the error code, refer to the user’s manual IM 12J05D02-01E for 

DO402G Dissolved Oxygen Converter, IM 12A01F03-01EN for FLXA402 4-Wire Converter 

Operation of Converter.

 

8th Edition : July 11, 2023

Содержание DO70G

Страница 1: ...User s Manual Model DO70G Optical Dissolved Oxygen Sensor IM 12J05D04 01E IM 12J05D04 01E 8th Edition...

Страница 2: ...mation on the converter connecting holders with the DO70G and how to connect sensor cables Note that the instruction manuals listed in the following table are for the associated equipment used with th...

Страница 3: ...perator must take to avoid such risks CAUTION This symbol indicates that the operator must refer to the instructions in this manual in order to prevent the instrument hardware or software from being d...

Страница 4: ...s of warranty period Failure of components which are out of scope of warranty stated in instruction manual Failure caused by usage of software hardware or auxiliary equipment which Yokogawa did not su...

Страница 5: ...Blank Page...

Страница 6: ...2 Installing the Sensor and Connecting the Cable 2 1 2 1 Preparing for Installation 2 1 2 2 Installation 2 2 2 2 1 Selecting the Measurement Point 2 2 2 2 2 Selecting the Sensor Holder 2 3 2 2 3 Insta...

Страница 7: ...Blank Page...

Страница 8: ...ng ball or vertical floating ball holders In the submersion type holder the sensor can be automatically cleaned with jet sprayed water or air at regular intervals By selecting a suitable holder and cl...

Страница 9: ...shape pin Weight Sensor Approx 0 1 kg Cable For E Approx 0 1 kg m 0 1 kg For D Approx 0 8 kg m 0 8 kg Power supply DO402G from DOX10 Power Supply Unit FLXA402 from FLXA402 The DOX10 may not be used de...

Страница 10: ...r Option Code If you order only DO70G please make a separate order for the adaptor by Part No in the table below However if you are already using DOX8HS PB350G PB360G and DO70G you can use the existin...

Страница 11: ...03E 05E 10E No cable type Approx 70 Approx 90 Approx 100 1500 L 3 m 5 m 10 m L 5 m 10 m 15 m 20 m Unit mm 12 7 Cable length code 05D 10D 15D 20D l Adaptor S3 PP Adaptor FPP 23 Pg13 5 DIN40430 30 5 26...

Страница 12: ...the cable connector insert the sensor connector into the cable connector Note A weak pressure is sufficient for inserting it If a strong pressure is needed to insert it the connection position will b...

Страница 13: ...r The assembly is submerged to the optimum point to obtain precise measurements This section explains how to select the measurement point and sensor holder 2 2 1 Selecting the Measurement Point Normal...

Страница 14: ...is immune to large variations in the liquid level Since the wet part is smooth and less susceptible to catching flocs the sensor does not trap rubbish that mixes in with the SUM Related description S...

Страница 15: ...r CAUTION Do not contaminate or wet the tip of the sensor cable when assembling the sensor If you will not begin wiring the sensor cable immediately take the protective measures necessary to prevent t...

Страница 16: ...solved Oxygen Converter user s manual IM 12J05D02 01E For information on sensor cable to connect DO70G to DO402G refer to the user s manual IM 12J05D02 01E for DO402G Dissolved Oxygen Converter instru...

Страница 17: ...Blank Page...

Страница 18: ...t may occur DO402G Dissolved Oxygen Converter or FLXA402 4 Wire Converter detects these failures If a sensor error code appears on the display follow one of the procedures below 1 Clean the sensor sen...

Страница 19: ...prepare 1 liter of the calibration solution for zero adjustment O Ring EPDM K9691KB For Adaptor Mounting Adaptor K9148NA K9148NB K9679CA For submersion type holder stainless steel S3 For submersion t...

Страница 20: ...l Added a maintenance part on the list page 4 2 Deleted CMPL Jul 2017 3rd Edition Added Trademark Notices Product Disposal and Model name pages i ii 1 5 Dec 2015 2nd Edition P1 3 P3 5 Added coments of...

Страница 21: ...Blank Page...

Отзывы: