3-14
IM DLM4038-03EN
3.6 Resetting the DLM4000 to Its Factory Default
Settings
Procedure
Resetting the DLM4000 to Its Factory Default Settings (DEFAULT
SETUP)
1.
Press the
DEFAULT SETUP
key.
The DLM4000 is reset to its factory default settings.
An Undo menu item appears.
Undoing the Reset Operation (Undo)
2.
Press the
Undo
soft key.
The previous settings are restored.
Explanation
You can reset the DLM4000 settings to their factory default values. This feature is useful when you
want to cancel all of the settings that you have entered or when you want to redo measurement from
scratch.
Settings That Cannot Be Reset to Their Factory Default Values
• Date and time settings
• Communication settings
• Language settings
• Measured value font size setting
Undoing the Reset Operation
If you reset the settings by mistake, you can press the Undo soft key to restore the previous settings.
However, you cannot undo the reset operation if you clear the Undo menu item by switching to a
different setup menu or pressing the ESC key.
To Reset All Settings to Their Factory Default Values
While holding down the RESET key ( ), turn the power switch on. All settings are reset to their
factory default values except the date and time settings (the display on/off setting will be reset) and the
setup data stored in internal memory.