15-4
IM DLM3054-01EN
When I
2
C bus signals are being analyzed, the display of the decoded address pattern is affected by whether the
R/W bit is on or off.
0 1 0 1 0 0 1 1
2
9
5
3
In the binary display, the address and R/W
bit are both in binary (01010011).
In the hexadecimal display when Include R/W
is off, the address is in hexadecimal (0x29),
and the R/W bit is in binary (1).
In the hexadecimal display when Include R/W
is on, the address and R/W bit are both in
hexadecimal (0x53).
1
R/W bit
7-bit Address Example
Address
List Display (List)
You can display a list of decoded results. You can list the decoded results of four serial bus signals. You can also
display the waveform of a specified analysis number in the zoom window.
Listing is not possible when the source is a user-defined serial bus signal.
Turning Zoom Linking On and Off (Zoom Link)
When Zoom Link is set to ON, the analysis number and zoom position settings are linked. When Zoom Link is
set to OFF, the analysis number and zoom position settings are not linked. The default setting is ON.
List Size (List Size)
Select the list size and display position from one of the settings below.
• Full Screen: Displays the list in full-screen mode
• Half(Upper): Displays the list in the top half of the screen
• Half(Lower): Displays the list in the bottom half of the screen
Detailed Display (Detail)
If the bus type is UART(When Grouping is on), I
2
C, or SPI, the instrument displays all of the data for the
specified analysis number in hexadecimal notation and ASCII codes. If you show the detailed display, you can
view the data from the 17th byte.
Analysis Number (List No.)
The line on the list that corresponds to the number you select is highlighted. When Zoom Link is set to ON,
the zoom window displays the waveform expanded with the start of the frame that corresponds to the selected
analysis number at the center of the window.
Trend Display (Trend)
You can display the trend of decoded SENT signal data. For each analysis of serial bus signals (Serial Bus1 (S1),
Serial Bus2 (S2), Serial Bus3 (S3), Serial Bus4 (S4)), up to four trends can be displayed.
Saving the Analysis Results
You can save the analysis results (the data list of the decoded results) in CSV format to a storage device. The
extension is .csv.
For details on saving the serial bus analysis results, see “Serial Bus Analysis Results (Serial Bus)” in chapter 20,
“Saving and Loading Data.”
You cannot save the results of analyzing a user-defined serial bus signal (User Define).
15 Analyzing and Searching Serial Bus Signals