7-1
IM DLM3054-01EN
7 Display
Types of Windows
This instrument has the following types of windows.
VT waveform display window
• Main window
Displays normal waveforms, which are not magnified
• Zoom1 window
Displays zoomed waveforms according to the settings specified using the ZOOM1 key.
• Zoom2 window
Displays zoomed waveforms according to the settings specified using the ZOOM2 key.
XY window
• XY1 window
Displays the XY1 waveform according to the settings specified using the X-Y key.
• XY2 window
1
Displays the XY2 waveform according to the settings specified using the X-Y key.
FFT window
2
Displays the FFT1 and FF2
1
waveforms according to the settings specified by the FFT key.
Trend/Histogram window
2, 3
Displays Trend1, Trend2, Hist1, and Hist2 according to the statistical processing settings specified using the
MEASURE key.
1 XY2 and FFT2 waveforms are only available on 4-channel models.
2 The FFT window and the Trend/Histogram window are the same.
3 Only Trend1 or Hist1 can be displayed on 2-channel models. Up to two items can be displayed on 4-channel
models.
Display Example
<FFT, Trend,
Histogram>
<XY2>
<XY1>
<Zoom1>
<Main>
<Zoom2>
Display Format (Format)
You can evenly divide the VT waveform display window so that you can easily view input waveforms and
computed waveforms. You can divide the window in the following ways.
Auto,
1
Single (one area), Dual (two areas), Triad (three areas), Quad (four areas), Hexa (six areas),
2
Octal (eight
areas)
2
1 The number of areas is automatically selected based on the number of displayed waveforms.
2 Only on 4-channel models
The number of displayed points per area varies depending on how many areas the window is divided into.
The vertical resolution does not change even if the number of displayed points is different.
The number of displayed points when only the Main window is displayed is as follows:
Single: 640 points, Dual: 320 points, Triad: 208 points, Quad: 160 points, Hexa: 104 points, Octal: 80 points