9.9 Counting Rotations
Procedure
1.
Carry out steps 1 to 5 in section 9.1 to set the operator to Rotary Count.
Setting the Count Conditions
2.
Press the
Setup
soft key.
3.
Press the
Source1
soft key. A menu used to select the waveform to be computed
appears. Press the soft key corresponding to the source waveform.
To select REF1 to REF4, first switch to
REF1-REF4
by pressing the corresponding soft key.
When CH1 to CH4 on the front panel are set, the waveform to be computed is fixed to the
channel that is set. A menu used to select the waveform to be computed don’t appears.
4.
Use the
rotary knob
to set the determination level of phase A.
5.
Press the
Source2
soft key twice to display the Source2 menu.
6.
Press any of the
CH1
to
CH4
and
REF1-REF4
soft keys to select Source2.
To select REF1 to REF4, first switch to
REF1-REF4
by pressing the corresponding soft key.
7.
Use the
rotary knob
to set the determination level of phase B.
8.
Press the
Initial Point
soft key.
9.
Use the
rotary knob
to set the count start point. Or, press the
Jump to
soft key
to show a preset menu, and set the count start point.
Preset
menu
Setting the Unit and Display Range
Carry out steps 6 to 13 in section 9.1.