5-11
IM 701310-01E
Vertical and Horizontal
Axes
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
5.9 Using the Auto Scale Function
Procedure
1.
Press one of the
CH1
to
CH4
keys to select the channel.
2.
Press the
Select
soft key to select INPUT or MATH.
For INPUT:
3.
Press the
Auto Scale EXEC
soft key.
For MATH:
3.
Press the
Ranging
soft key to select Auto.
Explanation
Auto Scale can be set for each channel.
The following are set in the following ways when Auto Scale EXEC is used:
V/div
Displayed so that the entire amplitude of the waveform can be seen without
changing the vertical position.
Offset
0 V when the input coupling is AC.
Center = (Max – Min)/2 when the input coupling is DC.
Trig Level
DC offset position.