21-9
IM DL850E-01EN
21 Other Features
Intensity (Intensity)
You can set the intensities of the grid (Grid), cursor (Cursor), and marker (Marker) to values within the range of 1
to 8.
Key and Knob Setup (Key/Knob Setup)
Turning On or Off the Click Sound (Click Sound)
You can turn on or off the click sound that is generated when you operate the jog shuttle.
START/STOP Key Response Time (START/STOP Response Time)
You can set the response time of the START/STOP key to instant (Quick) or 1 s or more (> 1sec).
Key Lock (Key Protect)
You can lock the operation keys to prevent unintentional changes to the current state of the DL850E/DL850EV.
• Type (Type)
Select whether to lock all the keys (ALL) or to lock all keys except the START/STOP key (Except START/
STOP).
• Release Method (Release Type)
Select whether to release the key lock by pressing the KEY PROTECT key (Key) or by entering a password
(Password).
• Password (Password)
Specify the password to use to release the key lock. Specify the password using up to eight alphanumeric
characters. If you forget the password, you can release the key lock by turning the DL850E/DL850EV on
while holding down the RESET key. Note that all settings will be initialized when you do this.
Setting the Cursor Read Mode and Data Save Destination (Analysis Setup)
Cursor Read Mode (Cursor Read Mode)
You can select whether to perform cursor measurements on P-P compressed display data or the data that has
been acquired in the acquisition memory.
• Display data (Display)
Cursor measurements are performed on the display data.
• Acquisition (ACQ)
Cursor measurements are performed on sampled data in acquisition memory.
Setting the Destination That Data Is Saved to upon Action Execution (Action Folder Mode)
You can select the destination that data is saved to when actions are executed.
• ON: Data is saved to the folder that is automatically created with the date.
If the number of files in the save destination folder exceeds 1000, a new folder is automatically created with
the date and an incremented sequence number (000 to 999) as its name, and the data continues to be saved
in the new folder.
• OFF: Data is saved in the folder that you have specified.
• The maximum number of files that can be saved to a single folder is 1000. Make sure that there are no
files in the destination folder before you execute an action.
• If the measurement count (Acquisition Count) is set to a number greater than 1000, the measurement
cannot be started.
Setting the DC Offset and Gain Adjustment (Input Setup)
Turning DC Offset and Gain Adjustment On and Off (DC Offset & Gain Adjust)
Select whether to use offset adjustment or gain adjustment (ON/OFF).