11-4
IM DL850E-01EN
Scaling Mode (Scaling Mode)
Set the method used to set the vertical display range of computed waveforms to one of the following options.
• Auto: The upper and lower limits are set automatically.
• Manual: The upper and lower limits must be set manually.
Upper and Lower Limits (Upper/Lower)
Set the upper and lower limits when Scaling Mode is set to Manual.
The selectable range is −9.9999E+30 to 9.9999E+30.
Start Point and End Point (Start Point/End Point)
Set the range of computation by specifying a computation start and end point. The default settings are −5 div
and +5 div.
The maximum range from the computation start point to the computation end point varies as indicated below
depending on the number of computations.
• One computation: Up to 1 Mpoint
• Two computations: Up to 500 kpoint
• Three or four computations: Up to 250 kpoint
• Five to eight computations: Up to 125 kpoint
The computation range points are similar to the settable range of the cursor display position in cursor
measurement.
For details, see “Selectable Range of Cursor Positions.”
An icon (
) appears in the center of the top of the screen when computations are being executed.
User-Defined Computation (Optional)
You can perform user-defined computation on models with the /G2 option.
Operators and Functions (Operation)
Select User Define.
Expression (Expression) - user-defined
Create an expression by combining computation source waveforms, variables, constants, functions, and
operators.
Computation Source Waveforms and Variables
You can use the following waveforms and variables.
Menu Item
Description
C1 to C16
Waveforms from CH1 to CH16
Cn_1 to Cn_16
• The waveforms of sub channels 1 to 16 on a 16-CH voltage input module or a 16-CH temperature/
voltage input module
• n = 1, 3, 5, 7, 9, 11, 13, 15
Cn_1 to Cn_16
• The waveforms of sub channels 1 to 16 on a CAN bus monitor module (You can configure up to 60
sub channels on modules whose firmware version is 1.20 and later.)
• The waveforms of sub channels 1 to 60 on a CAN & LIN bus monitor module.
• n = 13 to 16
• If the data type (Value Type) is set to Logic, you will not be able to select it.
Cn_1 to Cn_8
Logic module waveforms
n is the channel number. Specify using the following syntax: Cn_1, upper limit, lower limit.
M1 to M7
Math waveforms
T
Total number of data points in the time direction
When computation is performed on a linearly scaled channel, the scaled values are used.
11 Computation