App-28
IM DL850-03EN
Appendix 6 High-Voltage 100 kS/s, 16-Bit
Isolation Module (with RMS) (701260)
Specifications
The specifications for the High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701260) are
provided below. This input module (701260) will not comply with safety standard EN61010-1 or
EN61010-2-030 after September 30, 2013.
Item
Specifications
Standard operating
conditions
Temperature:
Humidity:
23 ±5°C
20 to 80% RH
After a 30-minute warm-up and after calibration
Effective measurement
range
20 div (±10 div around 0 V, display range: 10 div, when Variable is OFF)
Number of input channels 2
Input coupling
AC, DC, GND, AC-RMS, and DC-RMS
Maximum sample rate
100 kS/s
Input format
Isolated unbalanced
Frequency characteristics
1
(−3 dB point when a sine
wave of amplitude ±3 div
is applied)
Waveform observation mode:
RMS observation mode:
DC to 40 kHz
DC, 40 Hz to 10 kHz
Voltage-axis sensitivity
setting
20 mV/div to 200 V/div (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or
less)
Combined with the 700929(10:1) or 701947 (100:1):
2
Combined with the 701954 (1:1):
5
Direct input or cable not complying with the safety standard:
8
1000 V (DC+ACpeak)
850 V (DC+ACpeak)
850 V (DC+ACpeak)
Maximum allowable
common mode voltage
(at a frequency of 1 kHz or
less)
Working voltage of safety standard
Combined with the 700929 or 701947 (across probe tip H and earth
3
):
(across probe tip L and earth
4
):
Combined with the 701954 (1:1)
(across tip H and earth
6
):
(across tip L and earth
7
):
1000 Vrms (CAT II)
400 Vrms (CAT II)
700 Vrms (CAT II)
400 Vrms (CAT II)
Direct input or cable not complying with the safety standard:
9
(across the input terminal, H or L, and earth)
30 Vrms (42 VDC+ACpeak)
Vertical (voltage) axis
accuracy DC accuracy
1
Waveform observation mode
DC accuracy
RMS observation mode
DC accuracy
AC accuracy (when a sine wave is input)
AC accuracy (when the crest factor is 2 or less)
AC accuracy (when the crest factor is 3 or less)
±(0.25% of 10 div)
±(1.0% of 10 div)
±(1.5% of 10 div)
±(2.0% of 10 div)
±(3.0% of 10 div)
At frequency of 40 Hz to 1 kH
At frequency of 40 Hz to 1 kH
At frequency of 40 Hz to 1 kH
Input connector
BNC connector (isolated type)
Input impedance
1 MΩ ± 1%, approx. 35 pF
−3 dB point when AC
coupled low frequency
attenuation point
1 Hz or less (0.1 Hz or less when using the 700929, 0.01 Hz or less when using the 701947)
Common mode rejection
ratio
80 dB (50/60 Hz) or more (Typical
10
)
Residual noise level
(Input section shorted)
±1 mV or ±0.02 div whichever is greater (Typical
10
)
Withstand voltage
3700 Vrms for 1 minute (across each terminal and earth) (60 Hz)
Allowable transient surge
voltage (instantaneous)
±5200 Vpeak (across each input terminal and earth)
Insulation resistance
500 VDC, 10 MΩ or more (across each input terminal and earth)
A/D conversion resolution 16 bit (2400 LSB/div)
Temperature coefficient
Zero point:
Gain:
±(0.02% of 10 div)/°C(Typical
10
)
±(0.02% of 10 div)/°C(Typical
10
)
Response time (only when
observing RMS)
Rising (0 to 90% of 10 div):
Falling (100 to 10% of 10 div):
100 ms (Typical
10
)
250 ms (Typical
10
)
Bandwidth limit
Select from OFF, 10 kHz,1 kHz, and 100 Hz
Cut-off characteristics: −12 dB/OCT (Typical
10
)
Probe attenuation setting Voltage probe:
Current probe:
1:1, 10:1, 100:1, 1000:1