2-9
IM 701450-51E
CAN Bus Signal Analysis Function
2
Setting the Combination Trigger
24.
Press the
Combination
soft key. The Combination menu appears.
25.
Press
CAN Only
,
CAN on Pattern
, or
CAN -> Pattern
soft key.
Select CAN Only to activate a trigger only on the trigger conditions of the CAN Bus
signal. Select CAN on Pattern to activate a trigger when the trigger conditions of the
CAN Bus signal are met while the trigger conditions of CH2 to CH8 (CH2 to CH4 on the
DL7440) are met. Select CAN -> Pattern to activate a trigger when the trigger
conditions of CH2 to CH8 (CH2 to CH4 on the DL7440) are met after the trigger
conditions of the CAN Bus signal are met.
When CAN Only Is Selected
26.
Press the
Set Pattern
soft key. The Set Pattern dialog box displays the
assignment conditions of the CAN Bus signal of CH1.
27.
Press
ESC
. The Set Pattern dialog box closes.
Proceed to step 32.
2.4 Setting CAN Bus Signal Acquisition Conditions