50
IM 701450-61E
Appendix 1
Setup Parameters That Are Changed
during the Execution of Auto Deskew
The settings of the following parameters are changed when auto deskew is executed.
Panel Key and Knob
Soft Key
Setting
CH1 to 6 (CH5 and CH6 can be used only on the DL7480): Channel
Channel with voltage probe connected (CH1, CH3, or CH5)
Display
ON
Position
–3.00 div
Coupling
DC1M
Ω
Offset
0 V
Linear Scale
OFF
Channel with current probe connected (CH2, CH4, or CH6)
Display
ON
Position
2.00 div
Coupling
DC1M
Ω
Offset
0 V
Linear Scale
OFF
V/DIV: Vertical axis
Channel with voltage probe connected
1 V/div
(2 V/div when probe attenuation is 1000:1)
Channel with current probe connected
20.0 mA/div
TIME/DIV: Time axis
20 ns/div
MODE: Trigger mode
When correction is executed and is successful
Normal (DL7400 in start condition)
When correction is executed but is unsuccessful
Single (DL7400 in stop condition)
SIMPLE: Simple trigger
When the trigger source is the channel that has the voltage probe
connected
Level
3.00 V
When the trigger source is the channel that has the current probe
connected
Level
–40.0 mA
Slope
Falling edge
POSITION: Trigger position
Position
50%
DELAY: Trigger delay
Delay
0.00 ns
ACQ: Waveform acquisition conditions
Record Length
10k
Mode
Normal
Repetitive
ON
Time Base
Int
MEASURE: Automated measurement of waveform parameters
Dual Area
OFF
1Cycle Mode
OFF
Time Range
–5.000 to 5.000 div
Delay Setup
Reference
Trig
Dist/Prox Mode
%