Trigger

User’s manual chapter 6, “Triggers”

There are two main categories of triggers: simple triggers and enhanced triggers.

Simple triggers

Edge trigger

If the signal input to the 
measurement input 
terminal passes through 
the specified trigger level 
(on the rising or falling 
edge, or both edges), the 
trigger activates.

External trigger

If the signal input to the 
EXT TRIG IN terminal 
passes through the 
specified trigger level (on 
the rising or falling edge, or 
both edges), the trigger 
activates.

Commercial power 
trigger

A trigger activates on the 
rising or falling of the 
waveform from the 
commercial power being 
supplied to the DL7400.

Enhanced triggers

A->B(N) trigger

A trigger activates the Nth time 
condition B becomes true after 
condition A becomes true. 

A Delay B trigger

After condition A becomes true, a 
trigger activates once condition B 
becomes true after a specified time 
has elapsed.

Pattern trigger

A trigger activates when all of the 
trigger conditions from several trigger 
sources* are met/not met.

Width trigger

A trigger activates depending on 
whether the conditions became true 
or not for a shorter or longer duration 
than a specified reference time.

OR trigger

A trigger activates when at least one 
of the trigger conditions from several 
trigger sources is met.

Window trigger

A trigger activates when a trigger 
source enters or exits a “window” 
specified by two voltage levels.

TV trigger

A trigger activates on the video signal 
being input to CH1.

Logic trigger

A trigger activates on a logic signal.

* A trigger source is a signal to which trigger activation conditions are applied.

History Memory

User’s manual section 10.1, “Displaying History Waveforms”

When measuring waveforms, numerical data is being loaded into the acquisition memory, and that data can be

viewed on the DL7400 screen as a waveform.  When taking continuous measurements, even if you stop

measurement when you see an abnormal waveform, the waveform usually scrolls off the screen by the time

measurement stops, preventing you from viewing it carefully.  But with the history memory function, you can display

past waveform data that has been loaded into the acquisition memory while measurement is stopped.  You can

displayed a specific waveform from among a maximum of 4096 “history waveforms.”  Also, you can search for history

waveforms that pass (or do not pass) through a specified search zone.  For details on this function, see section 10.2,

“Searching the History Waveforms Using Zones (History Search)” in the user’s manual.

History waveform

Selected Record No. 0

Selected Record No. –25

Specified search zone

Searches for waveforms 
that pass through the 
specified search zone.

Содержание DL7440

Страница 1: ...DL7440 DL7480 Digital Oscilloscope IM 701450 02E 2nd Edition...

Страница 2: ...y questions or find any errors please contact your nearest YOKOGAWA representative as listed on the back cover of this manual Copying or reproducing all or any part of the contents of this manual with...

Страница 3: ...y and Jog Shuttle Operation 10 Introduction to Main Functions 12 Working with the DL7400 17 Preparing for an Observation 17 Displaying a Waveform on the Screen 19 Changing Waveform Display Conditions...

Страница 4: ...ns Waveform Display Conditions Installing the Instrument Connecting the Power Supply and Turning ON OFF the Power Switch Connecting the Probe Initializing the Settings Performing Auto Setup Vertical A...

Страница 5: ...the current axis sensitivity is set when setting the current to voltage conversion ratio PRESET key Section 5 7 Displays a menu for automatically setting the probe attenuation current to voltage conv...

Страница 6: ...he record length acquisition mode interleave mode sampling rate and time base during waveform acquisition START STOP key Section 7 1 Starts stops waveform acquisition according to the trigger mode The...

Страница 7: ...ess backlight self test and remote control It can also be used to display settings and the system status options installed on the unit firmware version etc SNAP SHOT key Section 8 6 Displays a non upd...

Страница 8: ...e from the trigger position Label of the displayed waveform Section 8 9 Scale value Section 8 8 Comment The comment specified when printing or saving the screen image is displayed Trigger type Section...

Страница 9: ...lay frame Z2 display frame Record length for the Z1 display T div for Z2 T div for Z1 Record length for the main normal waveform display Record length for the Z2 display Screen When Displaying Logic W...

Страница 10: ...en Screen When Displaying X Y Waveforms User s manual section 8 5 Displaying X Y Waveform Main normal waveform display area X Y waveform display frame X axis scale Y axis scale Waveform assigned to X...

Страница 11: ...isplay the corresponding menu Note Selecting an item among displayed selections Each time you press the soft key the selection item changes The selected item is highlighted Selecting Items by displayi...

Страница 12: ...USB PERIPHERAL connector on the rear panel of the instrument you can use the mouse to perform the same operations available with the front panel keys Top menu Right click in the display waveform displ...

Страница 13: ...ference time OR trigger A trigger activates when at least one of the trigger conditions from several trigger sources is met Window trigger A trigger activates when a trigger source enters or exits a w...

Страница 14: ...tal axis and the signal level of another waveform assigned to the Y axis waveform for which display is ON and displays the relationship between the two You can observe an X Y waveform and the normal T...

Страница 15: ...me The zoom position automatically moves in the specified direction You can watch the zoomed waveform as it scrolls then stop it at any desired position Auto scroll Analyzing or Searching SPI Signals...

Страница 16: ...eters The following five results of statistical processing are performed on the same two measured items from automatic measurement of waveform parameters Max Min Avg Sdv Cnt count or the number of max...

Страница 17: ...een images Snapshot waveform Displayed in white Updated waveform Ethernet Communications User s manual chapter 13 Ethernet Communications Optional You can transmit data or control the DL7400 using the...

Страница 18: ...onnecting the power supply you must read the warnings in the user s manual listed in section 3 3 Connecting the Power Supply and Turning ON OFF the Power Switch 3 prong outlet Power cord accessory Pow...

Страница 19: ...e Probe Phase Correction Phase adjustment hole Functional ground terminal Probe compensation signal output terminal Probe For instructions on how to connect a logic probe see section 3 6 When actually...

Страница 20: ...rocedure are the following The date time setting communication settings stored settings the language setting and the SCSI ID setting To initialize all settings but the date time setting perform the pr...

Страница 21: ...default number of screen divisions is four Quad In this case we only want to observe channel 1 so we change the setting to Single 1 screen DISPLAY 2Open the selection menu then choose Single 1Press th...

Страница 22: ...ime Axis Setting from 200 s div to 100 ms div User s manual section 5 12 Setting T div The time axis setting is the setting for the amount of time per grid division If you slow increase the value of t...

Страница 23: ...l section 5 12 Setting T div The display reverts from roll mode back to update mode and five periods worth of the waveform is displayed 1Turn the TIME DIV knob clockwise to set the time axis setting t...

Страница 24: ...gger mode determines the conditions such as timing and the number of times at which the waveform display is updated If you execute auto setup the trigger mode is set to auto mode Five trigger modes ar...

Страница 25: ...nding soft key Trigger position mark Pre trigger portion Post trigger portion Trigger point 2Turn the jog shuttle counterclockwise to set Position to 30 0 The waveform shifts 2 div to the left Changin...

Страница 26: ...form Working with the DL7400 Measuring a Waveform This section explains how to use the vertical cursors to measure the voltage and period of the displayed waveform In addition to cursor measurement th...

Страница 27: ...or Z1 5Set Z2 Mag in the same manner If you activate the jog shuttle on both Z1 Mag and Z2 Mag you can set both zoom factors at the same time Zoom box Z2 Zoom box Z1 Main normal waveform display area...

Страница 28: ...ettings using this menu are not required by this operation guide but it may be useful to know that you can specify the print destination built in printer USB printer or network printer the printing fo...

Страница 29: ...ng Drive name of the storage medium See previous page These settings not required by this operation guide but you can enter them as needed 2Choose Waveform waveform data 3Display the Save menu 4Displa...

Отзывы: