5-7
5
IM 701830-01E
Vertical and Horizontal Axes
5.6
Selecting Probe Attenuation
≡
For a description of this function, refer to page 1-4.
≡
For High-Speed Isolation/High-Speed Module, set the attenuation according to the probe being
used. For High-Resolution, High-Voltage, Isolation/High-Resolution, Isolation Module, set the
attenuation to 1:1.
Function
Probe attenuation for each channel can be selected from the following.
1 : 1, 10 : 1, 100 : 1, 1000 : 1
Note
If an incorrect attenuation has been selected, V/div will also be incorrect. If you set the attenuation to 1:1
when using a 10:1 probe, for example, the displayed value for automatically measured amplitude will be 1/
10 the true value.
Operating Procedure
1.
Press one of the keys
CH1
to
CH16
to select the desired channel.
2.
Keep pressing the “
Probe
” soft key until the desired attenuation is selected.