2-5
IM 701730-17E
USB Interface
2
2.4 Setting up this Instrument (USB)
2.4
Setting up this Instrument (USB)
Procedure
Selecting the Communication Interface
1.
Press MISC. The MISC menu appears.
2.
Press the Remote Control soft key. The Remote Cntl menu appears.
3.
Press the Device soft key to select USB.
Note
Only the communications interfaces selected under Device are available. If commands are
sent using an unselected communications interface, the command will not be received.
Setting the ID Value
4.
Press the Own ID soft key.
5.
Turn the jog shuttle to set the ID value.
Explanation
Carry out the following settings when using a controller to set information that can be
specified through key operation on the instrument or when outputting setting parameters
or output waveform data to the controller.
Setting the ID Value
Set the ID value of the instrument within the following range:
1 to 127
You can connect multiple devices to a host controller on the USB. If the instrument is
connected to multiple devices in a single USB system, the ID value is used by the host
controller to identify each device. Therefore, unique ID values must be assigned to the
instrument within a single system.
Note
Do not change the ID value while using the USB.