Power Quality (Event) Threshold Settings
Press F1 (OFF/ON) to enable the threshold of each event to be entered.
Even if a threshold is specified, the event will not be detected if it is set
to OFF. When you set an event back to ON, the threshold set previously
appears.
Notes on Setting the Thresholds
Because the SWELL, DIP, and INT thresholds are set as percentages
of the nominal voltage, if the nominal voltage is changed, the threshold
voltages also change. The default value is the peak voltage (300%).
Because the InrushCurrent threshold is set as a percentage of the
current range, if the current range is changed, the threshold current also
changes. Keep these in mind.