<8. Troubleshooting>
8-5
IM 11R02A01-02E
5th Edition : Dec. 28, 2016-00
8.5 Computing Station
If the value is all normal, each input of the computing station shows the following normal values in
the table.
Mark of input data
Normal indication
NON-CORR.
WI %
*
A-PRESS
Pa
500 ± 50
G-PRESS
Pa
500 ± 50
SQT. DENSITY
Same as density meter
*:
When pre-heating circuit does not exist, 0 to 100% is almost in proportion to 0 to “the highest level WI value in the measurement
range.”
When preheating exists, about 20 to 100% is almost in proportion to 0 to “the highest level WI value of the measurement range.”.
The lamps of both alarm and fail of the computing station are lighted in the following case.
1) Lighting of the alarm lamp. (Yellow color)
It lights when the input or output signal is cut off. But in this case, the computing inside the
station keeps working. Change to Alarm display and examine the cause of the alarm.
2) Lighting of the fail lamp (Red color)
The lighting of the fail lamp means an occurrence of an abnormal trouble produced inside
the instrument. When a fail occurs, the analog output and status output reserve the value
just before the occurrence of the failure. Such hold output has a tendency of decreasing
gradually as time proceeds. When the fail lamp is on, the display of fail will be shown.
The alarm lamp lights for the abnormal operation of the pressure adjustment for the gas or
air line. The lighting when other than under the time of measurement is not abnormal.
8.6
Other Troubleshooting
Check and deal with measures mentioned in paragraph 8.1 to 8.5. As to the trouble not
mentioned in this manual, contact our service personnel and describe how the inspection goes,
and if you have any record..