4) Measure each of the output terminals on the terminal block with a voltmeter. Use the “zero
adjustment screw” on the differential pressure transmitter to adjust the voltage to 1 V.DC.
The terminal number assigned for gas is 15 and 16, and one for air is 17 and 18.
5) Use a flat-head screwdriver to make the adjustment. Turn the screw clockwise to increase
the output, and counterclockwise to decrease. The adjustment amount of the zero point
varies depending on how quickly the zero adjustment screw is turned. To make a fine
adjustment, turn the screw slowly, and to make a rough adjustment, turn the screw quickly.
6
When the adjustment is completed, connect the impulse lines while making sure there is no
leakage.
CAUTION
●
Be sure to make this adjustment every time the installation location of the panel is changed.
●
Power is supplied to the differential pressure transmitter from the calorie detector. To
operate the differential pressure transmitter, turn on the calorie detector switch of the switch
box and the POWER switch on the front panel of the calorie detector.
●
After the zero-point adjustment is completed, do not turn off the transmitter immediately. If
the power is turned off within 30 seconds after the completion of the adjustment, the value
will return to the last one before the adjustment takes place.
F0401.ai
Zero Adjustment
Screw
Figure 4.1 Differential Pressure Transmitter
4.7 Air Differential Pressure Adjustment
1) Switch the computing station to the Input/Output status display (refer to Figure 3.15) and
check X2 (A-PRESS).
2) Adjust V-16 so that the indicated value becomes “500.”
3) If the indicated value of the pressure gauge (2-3) has deviated from 20 kPa, set the value to
20 kPa using the pressure reducing valve (2-2) and make the adjustment described in (2)
again.
4.8
Pressure Adjustment of the Gas Line
The pressure adjustment of the gas line uses air.
n
For
town gas use
1) Close V-6 and connect the one touch coupler (2-5).
2) Open V-7 and V-8.