10-10
IM CA71-E
Step 9:
The 0 and
FS
symbols stop blinking, causing the calibrator
to return to the state discussed in step 4.
Using the measurement range setting rotary switch,
select the next range.
By repeating steps 4 to 8, you can adjust the measurement
function assigned to that range.
TIP
Press the
key while holding down the
key.
This key operation cancels the CAL mode.
You can use the same key operation to cancel the CAL mode during calibration,
before saving to memory.
10.4 Notes on the Adjustment of
Temperature Ranges - CAL71 only -
Adjusting the temperature measuring ranges of the thermocouple
function involves using special equipment to make reference junction
compensation adjustments.
For this reason, contact the vender from which you purchased
the instrument to perform this adjustment.
10.5 Post-adjustment Verification
When adjustment work is done, test the calibrator to ensure that
adjustments have been made correctly and adjustment setpoints
have been saved into memory correctly.
To do this test, turn off the calibrator once and turn it back on again.
Then, place the calibrator in normal source or measurement mode
and check the setpoints using the same calibration standard.
10.4 Notes on the Adjustment of Temperature Ranges - CAL71 only -