Connection Diagram (SOURCE, SIMULATE)
SIMULATE
(SINK)
Black
Red
Digital multimeter
(DMM)
SOURCE
Red
Black
I (A)
L
Digital multimeter
(DMM)
I (A)
L
H
L
DC voltage and
current generator
5.3 What to Check after Calibration
After you have finished calibrating the CA450, make sure (through inspection) that the
calibration has been performed correctly and that the calibration values have been saved to the
memory.
Inspection Method
After calibration has finished, turn off the CA450.
Then, turn on the CA0, and check the calibration in the normal output and measurement
modes (not calibration mode).
Use the reference device that you used for calibration to check the calibration. (Refer to the
calibration points.)