4.6.4 Setting the Slow Step Time
If you select slow step mode for current sweep output, you can select the hold time (in seconds)
of each step.
1) Press the LIGHT or SHIFT key to display “t.Int.”
HOLD
2) Use ▲ (RANGE) and ▼ (REL) to select 15s, 30s, 45s, or 60s.
3) Press the HOLD key to save the setting. (After “SEt” appears, “t.lnt” will appear again.)
) To return to normal measurement from setup mode, hold down the LIGHT key for
second or longer.
4.6.5 Returning the Settings to Their Factory Default Values
You can reset all the items that you have changed (set) in setup mode to their factory default
values.
1) Press the LIGHT or SHIFT key to display the screen for initialization, which shows
“dEF.”
HOLD
2) The settings return to their default values when you press the HOLD key.
(After “donE” appears, “dEF” will appear again.)
) To return to normal measurement from setup mode, hold down the LIGHT key for
second or longer.
Initial Settings (Default Values) for the Setting Items
Setting Item
Initial Setting
Buzzer sound
ON
SENSOR mode
Subdisplay
0.0
Main display
0.00
Unit
A
The current span of the 00 mA range for
DCmA measurement
0 to 00 mA
Slow step time setting
s