background image

User's

Manual

Safety Precautions

The handling precautions, component names, and specifications are covered in the separate Getting 

Started Guide (IM CA310-02EN).

Be sure to read the Getting Started Guide before using this instrument.

Basic Setup

1. 

Turn the instrument on, and set the rotary switch to 

SETTING

.

2. 

Use the left-most up and down keys to select the setting and the right-most up and down keys to 

switch the options.

3. 

Press OUTPUT ON/OFF to confirm the setting.

 

POWER

Selects the setting Selects an option

Confirms the settings

TYPE

Setting

Auto power-off

Backlight

Sweep time

Measure and

source terminal

RJ sensor

(banana terminal)

Burnout detection

Temperature

reference

Extended sensor

Initialization of

span settings

Initialization of

settings

Options

ON

*

/OFF

ON/OFF

*

/TIMER

15s

*

/30s/45s/60s

A

*

/B

ON

*

/OFF

ON

*

/OFF

ITS-90

*

/IPTS68

D

*

/G/PL2

Display

*: Default value

Auto power-off

  If this is set to ON when the instrument is running on batteries, the power automatically turns off 

if you do not operate the instrument for approx. 20 minutes. This is invalid when the instrument is 

running off of the AC adapter.

Backlight

  Turns the backlight on and off. If set to Timer, the backlight turns on for approx. 2 minutes every 

time you operate a key.

Sweep time

  When sweeping in steps, set the output duration of each step. When sweeping linearly, set the 

output duration for the value to change from 0% to 100% (or 100% to 0%).

Measure and Source Terminal

  Set whether to use the thermocouple mini plug terminal (A) or the banana terminal (B).

RJ Sensor

  Sets whether to turn the reference junction compensation on or off when measuring temperature or 

sourcing temperature using the banana terminal.

Burnout

  Set whether to detect thermocouple burnouts.

Temperature Reference

  Set the temperature reference to ITS-90 or IPTS-68.

Extended Sensor

  Set the sensor to be assigned to the EXTRA range.

Initialization of Span Settings

  Initializes the span settings.

Initialization of Settings

  Initializes settings.

Using the RJ Sensor

This instrument has a built-in RJ sensor for reference junction compensation. In addition, you can also 

use an external RJ sensor (sold separately) to perform junction compensation based on an external 

contact.

Thermocouple Mini Plug Terminal

  The internal reference junction compensation is always enabled. External reference junction 

compensation using an external RJ sensor is not possible.

Banana Terminal

  If an external RJ sensor is not connected, the internal reference junction compensation is enabled. 

If it is connected, external reference junction compensation is enabled. To perform reference 

junction compensation with the banana terminal, set the RJ sensor in the basic setup (mentioned 

above) to ON.

Note

•  You cannot turn off the RJ sensor when the thermocouple mini plug terminal is in use.

•  If you are not using external reference junction compensation, be sure to remove the external 

RJ sensor.

This manual explains how to use the TC Calibrator.

Keep this manual in a safe place for quick reference in the event that a question arises.

Measurement

1. 

Set the CA310’s measure/source switch to 

MEASURE

.

2. 

Follow the instructions in the Getting Started Guide to connect the CA310 to the device to be 

measured using the included lead cable.

3. 

Use the rotary switch to set the range.

4. 

Press 

TYPE

 to select 

TYPE1 or TYPE2.

Note

•  When the range is set to EXTRA, the sensor selected for EXTRA in the basic setup is selected. 

  When the range is set to 90 mV, voltage ranging from -11.000 mV to 99.999 mV can be 

measured.

•  The included binding post can also be attached to the banana terminal.

Sourcing

1. 

Set the CA310’s measure/source switch to 

SOURCE

.

 

ON appears on the screen, and the output is turned on.

2. 

Use the rotary switch to set the range.

3. 

Press 

TYPE

 to select TYPE1 or TYPE2.

Note

When the range is set to EXTRA, the sensor selected for EXTRA in the basic setup is selected. 

When the range is set to 90 mV, voltage ranging from -11.000 mV to 99.999 mV can be sourced. It 

can be used to calibrate a thermocouple whose thermoelectromotive force falls within this range.

4. 

Follow the instructions in the Getting Started Guide to connect the CA310 to the target device 

using the included lead cable.

5. 

Follow the instructions for common operations in the Getting Started Guide, and set the output 

value with the ▼ and ▲ keys. The number of counts that change when the ▼ or ▲ key is 

pressed varies depending on the range.

Note

•  The CA310 is calibrated excluding the voltage dip in the lead cable.

  Load current readings include an error resulting from the lead cable resistance.

•  The included binding post can also be attached to the banana terminal.

Changing the Output Mode

1. 

Refer to “Keys” in the Getting Started Guide, and set the output mode.

 

Normal output (no display): The set value is output. The output value can be varied using the ▼ 

and ▲ keys.

 

Span check (

): In addition to normal output, the output can be varied in units of 25% of the 

span.

 

Step sweep( ): The output is varied automatically in steps of 25% of the span.

 

Linear sweep ( ): The output is varied linearly from 0% to 100%.

 

Press simultaneously to set the output mode.

Output setting for span check

Increase or 

decrease by 

1 count

Increase or decrease 

by 100 counts

Increase or decrease in 

25% steps of span

Set to 100% of 

span

Set to 0% of 

span

Pressing this when sweeping is 

paused sets the output to the 

span 0% value.

Sweeping

  The output value is varied automatically between 0% and 100% of the specified span. There are 

two sweep modes: step and linear.

 

Time

Output

0%

100%

Sweep time

Time

Output

0%

100%

Sweep time

Step sweep

Linear sweep

1. 

Set the 0% and 100% of the span, which is the sweep range.

 

Set the span 0% value using the ▼ and ▲ keys and then press simultaneously the first and the 

second ▼ keys from the right. SET is displayed, and 0% is set.

 

Likewise, set the span 100% value using the ▼ and ▲ keys and then press simultaneously the 

first and the second ▼ keys from the right.

 

For details on key operation, see “Keys” in the Getting Started Guide.

2. 

Set the output mode to step sweep or linear sweep. The ON indicator blinks.

3. 

Press 

TYPE

. Sweeping will start. The ON indicator changes to solid.

 

Pressing 

TYPE

 again pauses sweeping and causes the ON indicator to blink. The displayed 

value continues to be output while sweeping is paused.

 

Pressing 

TYPE

 yet again resumes sweeping from the last output value.

 

When sweeping is paused, pressing the left-most ▲ key sets the output to the span 0% value.

 

Changing the Sub Display

  The temperature shown on the main display can be displayed using a value converted into 

thermoelectromotive force or a percentage of the span. The temperature measurement of the RJ 

sensor is also displayed. This is invalid when the range is 90 mV.

 

Press simultaneously to set the sub display.

Thermoelectromotive force

Percentage of span

Temperature measurement of RJ sensor

CA320
TC Calibrator
User’s Manual

IM CA320-01Z2 

1st Edition

1st Edition: November 2015 (YMI)
All Rights Reserved, Copyright © 2015, Yokogawa Meters & Instruments Corporation
Printed in Japan

IM CA320-01Z2