4-16
IM CA150E
4.7 Source Frequency and Pulse (PULSE) Signals
4.7.1 Source a Continuous Pulse Train
1
Connect the terminals.
2
Use the FUNCTION key on the SOURCE side to select
frequency and pulse (PULSE). ( PULSE appears.)
3
Use the RANGE key to select a frequency setting mode.
(Default setting: 100 Hz)
(100 Hz, 1000 Hz, 10 kHz, 50 kHz, 1000 CPM)
Use the ▲ ▼ output value setting keys to set the frequency.
Each time a key is pressed the value is incremented or
decremented by one count.
If you attempt to increment or decrement the value 9 or 0,
the digit moves up or down one place.
A value continues to change when you hold down the key.
Press the CLEAR key to restore the frequency setting of
each range to its default setting.
4
Press the PULSE SET key once to switch to amplitude setting mode.
Use the ▲ ▼ output value setting keys to set the voltage.
Press the CLEAR key to restore the setting to 0.10000 V.
5
Press the PULSE SET key again to switch to pulse number
setting mode.
(Default setting: Cont, continuous pulse)
For generating a continuous frequency, no setting is necessary.
For generating a pulse cycle, set the pulse number.
6
Press the PULSE SET key again to return back to
frequency setting mode.
7
Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.)
8
Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
Tip
When frequency or pulse (PULSE) is selected using FUNCTION,
the sweep functions (step, linear, and program) and
the divided output (n/m) function cannot be used.