IM CA12E-02E <Page 1>

To return to the previous step:

<9> To return to the previous step without fixing the setting, press and hold the 

 input  

cancellation key for about 1 second.

 

Note that this is possible only before writing to the EEPROM.

To return to the nomal operation mode:

<10> Simultaneously press and hold the  

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds, or press 

 

the [POWER] key to turn off the power once and then press it again to turn it back on.

■ Calibration Precautions

(1) Calibration for 400 Ω generation

Set resistance-measuring current to 1 mA. 

 

Check the specifications of the digital multimeter (DMM) to be used as the standard.

Example:

 

 

Resistance-measuring current is 1 mA for the 2000 Ω range of Model 7561 (Yokogawa).

 

Connect the CA12E calibrator to the standard in a four-wire connection configuration 

 

 

for calibration as shown below:

3W

Lo Lo Hi

CA12E

DMM

2000 Ω range

H

L

Input terminal Sensing terminal

Standard

(2) Calibration for 400 Ω measurement

  Connect the CA12E calibrator to the standard in a three-wire connection configuration 

 

  for calibration as shown below:

3W

Lo Lo Hi

CA12E

Standard

Decade

resistance box

400 

Ω

(3) Calibration for internal RJ compensation

Because this calibration requires special equipment (a K-type thermocouple and 0°C 

 

reference temperature chamber), contact the sales agent or office where the instrument 

 

was purchased, or our office.

■ Assignment of Keys for Calibration

When the CA12E calibrator is in the calibration mode, keys are assigned as specified here.

1

2

3

4

CAL

ON

SOURCE

0

Input cancellation key

Input determination

(ENTER) key

Digit to adjusted by 

the adjustment keys 

Mode switching 

(calibration or nomal) 

(Press and hold two keys 

simultaneously for 2 seconds.) 

Adjustment keys

(six keys) 

(For about 1 second)

(For about 1 second)

■ Calibration the Measurement Feature

Operation procedure:

<1> Warm up the standard.

<2> Before turning on the power of the CA12E calibrator, connect it to the standard.

 

Be sure to remove the external RJ sensor.

<3> Turn on the power of the instrument.

<4> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds to enter 

 

the calibration mode.

<5> Select the measurement range to calibrate using the MEASURE/SOURCE 

 

selection switch and range selection rotary switch.

 

“CAL” and “MEASURE” appear and “FS” blinks on the display unit.

 

(If a value nearly equivalent to full scale has been input, a measured value and 

 

“FS” appear.)

<6> Set up the standard in order to input the full scale value to the instrument.

 

Wait until the reading stabilizes, then press and hold the   input determination 

 

(ENTER) key for about 1 second to fix the setting.

<7> After fixing the setting, “0” and “FS” indications on the display unit start blinking. 

 

Re-press and hold the     input determination (ENTER) key for about 1 second 

 

to write the calibration coefficients to the EEPROM of the instrument. 

 

(This overwrites the previous calibration coefficients.)

 

When this is complete, the instrument returns to the status in Step 5.

<8> Repeat Steps 5 to 7 for each range to be calibrated.

To return to the previous step:

<9> To return to the previous step without fixing the setting, press and hold the 

 input  

cancellation key for about 1 second.

 

Note that this is possible only before writing to the EEPROM.

To return to the nomal operation mode:

<10> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds, or press 

 

the [POWER] key to turn off the power once and then press it again to turn it back on.

Safety Precautions

The following manual, including this manual, are provided as manuals for the CA12E.

 

Please be sure to read all manuals.

 

IM CA12E-03EN (Getting Started Guide: For Safety)

 

IM CA12E-01E  (User's Manual - 1 - : Generation, Measurement, and Other features)

 

IM CA12E-02E  (User's Manual - 2 - : Calibration procedure and Specification): This manual

8.  Calibration Procedure

■ Calibration Procedure

To maintain a high level of accuracy, it is recommended that the CA12E HANDY CAL

 

 be calibrated

 

annually. 

 

The “Selecting the Standards” section below presents calibration methods using the 

 

recommended standards.

■ Selecting the Standards

<Source feature>

Items to be 

 

calibrated

Names of 

 

standards

Range

Measuring  

range

Accuracy

Remarks

DCV

Digital 

 

multimeter
(DMM)

100 mV

Max. 110 mV

±(0.005%+5 μV)

Model 7561
(YOKOGAWA) 

 

or equivalent

Ω

Digital 

 

multimeter

400 Ω

Max. 440 Ω
Measuring current: 

 

1 mA

±(0.005%+0.02 Ω)

<Measurement feature>

Items to be 

 

calibrated

Names of 

 

standards

Range

Measuring  

range

Accuracy

Remarks

DCV

Standard DC 

 

voltage generator

100 mV

100 mV

±0.01%

Model 2552
(YOKOGAWA) 

 

or equivalent

Ω

Decade resistance 
box

400 Ω

400 Ω

±0.01%

Model 279301
(YOKOGAWA) 

 

or equivalent

■ Environmental Conditions for Calibration

   Ambient temperature: 

23 ± 1°C

   Relative humidity: 

45 to 75% RH

   Warm-up:     

Warm-up time as specified for the standard

■ Calibration Points

• The calibration points are as specified in the following tables.

• It is possible to independently select the necessary range to be recalibrated.

• Always calibrate the zero (0) point and full scale (FS) as a pair for generation.

 <Source>      

 

 

<Measurement>

Calibration points

Standard Value *1

Calibration points

Standard Value *2

100 mV

0

0 mV

100 mV

FS

100 mV

FS

100 mV

400 Ω

0

0 Ω

400 Ω

FS

400 Ω

FS

400 Ω

*1:

Make adjustments to CA12E according 

 

to the reading of the standard 

 

(CA12E output Value) as specified in 

 

the table.

*2:

Set the value to the standard as 

 

specified in the table.

■ Calibration the Generation Feature

Operation procedure:

<1> Warm up the standard.

<2> Before turning on the power of the CA12E calibrator, connect it to the standard.

 

Be sure to remove the external RJ sensor.

<3> Turn on the power of the instrument.

<4> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds to enter 

 

the calibration mode.

<5> Select the generation range to calibrate using the MEASURE/SOURCE selection 

 

switch and range selection rotary switch. The display unit shows “CAL,” “SOURCE,” 

 

“ON,” “0,” and the lower limit.

<6> Read the output value of the CA12E using the standard (digital multimeter), and using 

 

the 

 adjustment keys adjust the CA12E so that the output value is set to the offset 

 

value. Then press and hold the   input determination (ENTER) key for about 1 second 

 

to fix the setting. 

 

After fixing the setting, the display unit reading changes to “CAL,” “SOURCE,” “ON,” 

 

“FS,” and a full scale of the range.

<7> Read the output value of the CA12E using the standard (digital multimeter), and using 

 

the  

 adjustment keys adjust the CA12E so that the output value is set to the full 

 

scale value. Then press and hold the   input determination (ENTER) key for about 

 

1 second to fix the setting.

 

After fixing the setting, the display unit shows “0 FS” blinking. Re-press and hold 

 

the 

 input determination (ENTER) key for about 1 second to write the calibration 

 

coefficients to the EEPROM of the instrument. 

 

(This overwrites the previous calibration coefficients.) 

 

When this is complete, the instrument returns to the status in Step 5.

<8> Repeat Steps 5 to 7 for each range to be calibrated.

User’s

Manual

CA12E

HANDY CAL

(Temperature Calibrator)

 

- 2 -

IM CA12E-02E

5th Edition: Dec. 2014

YMI

Printed in Korea

Keep this manual in a safe place so that

you can refer to it when necessary.