< 3. Installation >
3-5
IM 11M12D01-01E
Flow direction of the sample gas
Mount the protector so that the notch is
the downstream side of the gas flow.
Reference gas inlet
Calibration gas inlet
Unit: mm
F3-5E.ai
Probe top
1050,1550,2050
Gasket (Thickness 3.0)
Figure 3.4 Probe protector (for dust wear protect)
3.2 Installation of High Temperature Detector
(ZR22G-015)
3.2.1 Installation Location
This detector is used with the High Temperature Probe Adapter (Model ZO21P) when the temperature
of sample gas exceeds 700 °C, or when it is required due to maintenance spaces.
The following should be taken into consideration when installing the detector:
(1) Easy and safe access to the detector for checking and maintenance work.
(2) Ambient temperature of not more than 150 °C and the terminal box should not be exposed to
radiant heat.
(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The sample gas should satisfy the specifications described in Chapter 2.
3.2.2 Usage of the High Temperature Probe Adapter (ZO21P)
During analysis, the surface temperature of the probe adapter should be within the range from
more than the dew point of the sample gas and 300 °C or less to prevent ejector clogging, gasket
deterioration or bolt scoring.
Where the dew point of the sample gas is not known, keep within the range of more than 200 °C to
less than 300 °C.
The temperature shall be measured at the probe in the probe adapter and the surface of the blind
flange at the opposite side.
When the surface temperature is not within the above range, the following measures can be taken to
change the temperature.
<When the surface temperature exceeds 300 °C>
(1) When the furnace pressure is negative, lower the pressure setting to reduce induction flow
of the sample gas. Refer to Subsection 2.6.3, Ejector Assembly for High Temperature, for the
setting of induction flow. When you reduce induction flow, ensure that the ejector inducts air
when the furnace pressure fluctuates.
(2) When the furnace pressure is positive, close the needle valve for the sample gas outlet to
reduce the exhaust gas flow. Refer to Subsection 4.1.4, Piping to the High Temperature Probe
Adapter.