40
IM AQ6376-02EN
10 Specifications
Item
Specifications
Measurement wavelength range
1
1500 to 3400 nm
Span
1
0.5 nm to 1900 nm (entire wavelength range), 0 nm
Wavelength accuracy
1, 2, 5
±0.5 nm (entire wavelength range)
Wavelength repeatability
1, 2
±0.015 nm (1 minute)
Wavelength resolution setting
1, 2
0.1, 0.2, 0.5, 1 and 2 nm
Wavelength sampling resolution
1
0.003 nm minimum
Wavelength sampling points 101 to 50001, AUTO
Level sensitivity setting
NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3 (It is set to High
dynamic range mode (chop) when the sensitivity is set to HIGH1–HIGH3.)
Level sensitivity
2, 3, 6
-65 dBm (1500 to 2200 nm)
-55 dBm (2200 to 3200 nm)
-50 dBm (3200 to 3400 nm)
(measurement sensitivity: HIGH3)
Maximum input power
2, 3
+13 dBm (per channel, ful span)
Safe max. input power
2, 3
+20 dBm (total light input power)
Level accuracy
2, 3, 4
±1.0 dB (1550 nm, input level: -20 dBm, measurement sensitivity: HIGH1–HIGH3)
Leval linearity
2, 3
±0.2 dB (input level: -30 to +10 dBm, measurement sensitivity: HIGH1–HIGH3)
Dynamic range
1, 2
40 dB (±1 nm of peak wavelength)
55 dB (±2 nm of peak wavelength)
(resolution: 0.1 nm, measurement sensitivity: HIGH1–HIGH3)
Applicable fiber
SM (9.5/125 μm), GI (50/125 μm, 62.5/125 μm)
Optical connectors
For optical input, AQ9447(*) connector adapter (option) required.
For wavelength reference light source output, AQ9441(*) connector adapter (option) required.
(*): Connector types: FC or SC
Built-in calibration light source For alignment and wavelength calibration
Sweep time
1, 6, 7
0.5 s (NORM_AUTO), 1 s (NORMAL), 2 s (MID), 20 s (HIGH1)
Electrical
interfaces
GP-IB RS-232, Ethernet, USB, SVGA output, analog output port, trigger input port, trigger
output port
Remote control
8
GP-IB, RS-232, Ethernet (TCP/IP)
AQ6317 series compliant commands (IEEE488.1) and IEEE488.2
Purge gas I/O port
1/4” outer diameter nylon tube
Data storage
Internal storage
512 MB or more
Internal memory
64 traces, 64 programs, 3 templates
External storage
USB storage media (USB memory/HDD), format: FAT32
File types
CSV (text), binary, bitmap, TIFF
Display
9
10.4” color LCD (resolution: 800 x 600 pixels)
Environment conditions
Performance guarantee temperature: +18 to +28°C
Operating temperature: +5 to +35°C
Storage temperature: -10 to +50°C
Humidity: 20 to 80% RH (no condensation)
Location: indoor use, altitude: 2000 m or less
Warm-up
Warm-up time: 1 hour
After warm-up ends, alignment adjustment with the internal reference light source required.
Power requirement
100 to 240 VAC, 50/60 Hz, approximately 100 VA
Permitted supply voltage range 90 VAC to 264 VAC
External dimensions
Approximately 426 (W) x 221 (H) x 459 (D) mm (Note that this excludes the protector and
handle)
Mass
Approximately 23 kg
Recommended calibration period 1 year