10 Specifications
Item
Specifications
Measurement wavelength range
1
350 nm to 1750 nm
Span
1
0.5 nm to 1400 nm (entire wavelength range), 0 nm
Wavelength accuracy
1, 2, 5
±0.2 nm (entire wavelength range)
±0.05 nm (633 nm) (After wavelength calibration with a 633 nm HeNe laser)
±0.05 nm (1523 nm)
Wavelength repeatability
1, 2, 5
±0.015 nm (1 minute)
Wavelength resolution setting
1, 2
0.05 nm, 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm and 10 nm
Wavelength sampling resolution
1
0.002 nm minimum
Wavelength sampling points 101 to 100001, AUTO
High dynamic range mode SWITCH (Sensitivity: MID, HIGH1, HIGH2, HIGH3)
Level sensitivity setting
NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3
Level sensitivity
2, 3, 6
-70 dBm (400 nm to 900 nm)
-80 dBm (900 nm to 1600 nm)
(measurement sensitivity: HIGH3)
Safe max. input power
2, 3
+20 dBm (total light input power, 550 nm to 1750 nm)
+10 dBm (total light input power, 400 nm to 550 nm)
Level accuracy
2, 3, 4
±1.0 dB (1550 nm, input level: -20 dBm, measurement sensitivity: HIGH1–HIGH3)
Level linearity
2, 3
±0.2 dB (input level: -40 dBm to +0 dBm, measurement sensitivity: HIGH1–HIGH3)
Polarization dependency
2, 3, 4
±0.15 dB (1550 nm)
Dynamic range
1, 2
60 dB (633 nm, 1523 nm ±1 nm of peak wavelength) (resolution: 0.05 nm)
Applicable fiber
SM (9.5/125 μm), GI (50/125 μm, 62.5/125 μm), Large core size fibers (core diameter of up to
800 μm)
Optical connectors
For optical input, AQ9447(
□□
) connector adapter (option) required.
For wavelength calibration light source output, AQ9441(
□□
) connector adapter (option)
required.
(
□□
): Connector types: FC or SC
Built-in calibration light source For alignment and wavelength calibration
Sweep time
1, 6, 7
0.5 s (NORM_AUTO), 1 s (NORMAL), 2 s (MID), 5 s (HIGH1)
Electrical
interfaces
GP-IB RS-232, Ethernet, USB, SVGA output, analog output port, trigger input port, trigger
output port
Remote control
8
GP-IB, RS-232, Ethernet (TCP/IP)
AQ6317 series compliant commands (IEEE488.1) and IEEE488.2
Purge gas I/O port
1/4” outer diameter nylon tube
Data storage
Internal storage
512 MB or more
Internal memory
64 traces, 64 programs, 3 templates
External storage
USB storage media (USB memory/HDD), format: FAT32
File types
CSV (text), binary, bitmap, TIFF
Display
9
10.4” color LCD (resolution: 800 x 600 pixels)
Environment conditions
Optimal temperature: +18°C to +28°C
Operating temperature range: +5°C to +35°C
Storage temperature range: -10°C to +50°C
Ambient humidity: 20% to 80% RH (no condensation)
Warm-up
Warm-up time: 1 hour
After warm-up ends, alignment adjustment with the internal reference light source required.
Power requirement
100 to 240 VAC, 50/60 Hz, approximately 100 VA
External dimensions
Approximately 426 (W) x 221 (H) x 459 (D) mm (Note that this excludes the protector and
handle)
Mass
Approximately 19 kg
Recommended calibration period 1 year