3.2 Attaching the Connector Adapter
Attach the optional connector adapter before using the instrument.
WARNING
Always turn the power OFF before replacing the connector adapter. The instrument
has a built-in reference light source for wavelength calibration (when the built-in
light source specification is -L1), and infrared light is always being output from the
optical output connector. Never look into the optical output connector. Infrared light
entering the eyes can cause severe injury and loss of vision.
CAUTION
• As there may be dust adhering to calibration output, be sure to clean it before
attaching the connector adapter.
• Do not exhale or blow compressed air into the monochromator from the optical
input. Doing so may allow dust or other materials to enter the monochromator,
adversely affecting its optical performance. Also, if debris is adhering to the
optical components inside the monochromator when a strong light source is
input, the monochromator may be damaged.
• When attaching or removing the connector adapter, carefully insert it
perpendicularly to the ferrule so as not to damage the ferrule end.
• Moving the connector adapter to the right or left or inserting it forcefully can
damage the ferrule or the connector adapter.
A connector adapter is required for connecting the optical connector to the AQ6370D.
On products with the /FC, /SC, /RFC, or /RSC option, connector adapters come attached
to the optical input and calibration light source output on the AQ6370D front panel.
On products without these options, attach a connector adapter appropriate for the optical
connector.
Note
A different connector adapter is used for OPTICAL INPUT and CALIBRATION OUTPUT. Make
sure not to use the wrong connector adapter.
Attachment Procedure
1.
Confirm that the power is OFF.
2.
Open the optical connector cover at the front of the instrument.
3.
Clean the ferrule edge of the optical I/O section using a swab soaked with a small
amount of pure alcohol.
4.
Insert the connector adapter all the way in.
5.
Push the connector adapter’s lock lever down.
The adapter has been attached correctly if the groove in the lock lever interlocks with the
latch pin of the optical input/output section.