2.5 Connecting the DUT
WARNING
Do not look at the optical fiber laser light that you are measuring or point the laser
at another person’s eye. Doing so may cause eye damage or impair one’s health.
CAUTION
• Before connecting an optical fiber to the instrument, make sure that the start-up
initialization process has finished. If a strong light source is input during start-up,
the optical section can be damaged.
• Be sure to clean the tip of the optical fiber’s optical connector before connecting.
• As there may be dust adhering to calibration output, be sure to clean it before
connecting optical fiber.
• Do not try to forcefully attach the optical fiber’s optical connector with the plug
inserted at a slanted angle. Doing so may damage the instrument’s optical
connector’s components or the connector itself.
• Before connecting the input light, make sure that it does not exceed the
AQ6360’s maximum rated level. If input light exceeding the maximum rated level
is introduced, the optical section may be damaged.
• Do not exhale or blow compressed air into the monochromator from the optical
input. Doing so may allow dust or other materials to enter the monochromator,
adversely affecting its optical performance. Also, if debris is adhering to the
optical components inside the monochromator when a strong light source is
input, the monochromator may be damaged.
• Press the optical connector hard against the cleaning surface of the special
cleaner to clean it. If it is not pressed hard against the cleaning surface, it may
not be possible to properly clean the optical connector.
French
AVERTISSEMENT
Ne regardez pas directement la lumière du laser à fibre optique et ne pointez pas
le laser vers le yeux d’une tierce personne, pour ne pas provoquer de blessures ou
de dommages oculaires.