12-12
IM AQ1300-01EN
12.6 Configuring the CC Test
Procedure
CC Setup Screen
Follow the procedure in section 12.1 to display the E-OAM Test Setup screen.
Follow the procedure in section 12.3 to display the Select Test Mode screen, and then select
CC(Continuity Check). The following screen appears.
Set the interval
(100ms, 1s, 10s, 60s).
Set the maintenance domain information.
When the test setting is IEEE802.1ag:
Short MA Name Format
Short MA Name
When the test setting is ITU-T Y1731:
MEG ID Format
MEG ID
Set the network device (MEP) ID number
(0 to 8191).
Set the domain details.
Displayed when the test setting is IEEE802.1ag
Set the type of destination MAC address
(Unicast, Multicast).
Set the destination MAC address.
Valid when the destination MAC address
type is unicast
Refer to the MAC Address table.
Displayed when the destination MAC
address type is unicast
►
section 4.3
Select this check box to automatically
send an RDI when an LOC is detected.
Select this check box to send CCM frames.
Select this check
box to receive CCM
frames.
Set the transmission start timing of CCM frames
(Test Start, First Rx CCM).
Set the wait time.
Displayed when the transmission start
timing of CCM frames is set to Test Start
Select how to set the ID number of the
target network device (MEP) to receive
CCM frames from
(Designation, First CCM).
Set the ID number (0 to 8191).
Displayed when the method of setting the
ID number is set to Designation
Refer to the MEP ID table.
Displayed when the method of setting the
ID number is set to Designation