2-15
IM AQ1210-01EN
Setting Up the OTDR Feature
11
22
33
44
55
66
77
88
99
10
10
11
11
App
App
Index
Index
Setting the Cumulative Loss Type
5.
Tap
Cumul-Loss Type
to display the Cumul-Loss Type screen.
Select type
1, 2, or 3.
Note
With Type1, loss is measured from the measurement start point to the measurement point.
With Type2, loss is measured from the event marker (Y2) to the measurement point.
With Type3, loss is measured from the approximation line of the event marker (Y2) to the measurement
point.
For each type, the loss measurement range varies depending on whether the distance reference is set (top
illustration in the screen) or not set (bottom illustration in the screen). For details, see "Explanation" in this
section.
Setting Operation Restrictions
5.
Tap
Operation Lock Setup
to display the Operation Lock Setup screen.
Enter the administrator PIN (the default value is 0000).
Enter the new PIN again for confirmation.
Enter to operation
lock setup.
Enter the old
administrator PIN
(the default value
is 0000).
Enter the new
administrator PIN.
Confirms the input
2.3 Display (OTDR) Conditions