10.2 Using the Power Save Mode
Procedure
1.
Press
MENU
to display the MENU screen.
2.
Press
SETUP
to display the system setup screen.
3.
Tap
Power Save
to display the Power Save screen.
When using the USB-AC adapter
Set the LCD brightness (Bright, Normal, Power save, OFF).
When using the battery
Set the sleep function
(OFF, 1m, 5m, 10m, 30m).
Set the LCD brightness
(Bright, Normal, Power save, OFF).
Explanation
Setting the Power Save Mode
LCD Brightness
You can set the LCD brightness.
Bright
The screen is set bright. Use this setting in a bright environment. This setting uses a lot of
power. During battery operation, pay attention to the battery level.
Normal
Normal brightness.
Power save
Slightly darker the normal brightness. This setting can be used in a dark environment. The
battery lasts longer than the Normal setting.
OFF
The LCD backlight is turned off. The battery lasts longer, even longer than the Power Save
setting. When you press any key, the backlight turns on for approximately 10 seconds.
Sleep
If the instrument is left standing for a certain period with the power turned on, the power will be
automatically switch to sleep mode.
If the instrument is left standing for 2 hours in sleep mode, the power automatically turns off. This
does not apply during averaged measurement or real-time measurement.
OFF
The power will not be turned off.
1 min
If you do not operate the instrument for 1 minute, the power will be automatically switched to
sleep mode.
5 min
If you do not operate the instrument for 5 minutes, the power will be automatically switched to
sleep mode.
10 min
If you do not operate the instrument for 10 minutes, the power will be automatically switched to
sleep mode.
30 min
If you do not operate the instrument for 30 minutes, the power will be automatically switched to
sleep mode.