12.2 Error Messages
Error Messages
Messages may appear on the screen while you are using this instrument. This section explains the
meanings of the error messages and how to respond to them. You can display the error messages in
English or Japanese (see section 10.1). If servicing is necessary to solve the problem indicated by a
message, contact your nearest YOKOGAWA dealer.
Other communication-related error messages may also appear. These error messages are described
in the
Communication Interface User’s Manual
, IM This instrument-17EN.
Error in Execution
Code
Message
20
Network Option is not installed.
24
One or more conditions in this file are not supported by this product.
29
In remote control mode, all keys are locked except F1 key. Please hit F1 key to exit the remote control mode.
33
File operation is cancelled.
64
Failed to adjust LS power. Please check whether the patch fiber cable is properly connected between LS port and
OPM port.
65
Project Information doesn’t match between Master and Slave.
66
The project sent from master includes an unsuitable wavelength.
503
The measurement cannot be started during the file operation. Please either interrupt the file operation or wait until
the end of the file operation.
505
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
506
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
507
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
508
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
509
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
512
ZERO SET ERROR
513
ZERO SET ERROR
514
Exceeding limit. It may cause damage of the instrument. Please disconnect the plug.
515
ZERO SET ERROR
601
Cannot save onto the media. This media is for read-only.
602, 603
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
604
Storage media has been disconnected while the media is being accessed.
605
Same file name or folder name exists.
606, 607
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
608
Invalid file name or folder name
609
Invalid file format
610, 611
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
612
Invalid path name
613
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
614
Unknown file or folder
615
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
616
Cannot save onto the media. This media is for read-only.
617 to 619 Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
620
Free space is not enough
621
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
622
Unknown folder
623
Folder is not empty.
624 to 628 Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.