5.1 Producing Measurement Light
WARNING
• While this instrument is producing light, do not remove the optical fiber cable, because light
is emitted from the light source port. Visual impairment may occur if the light enters the
eye.
• Close the covers of any light source ports that do not have optical fiber cables connected
to them. On models with two or more light source ports, visual impairment may occur if light
that is mistakenly emitted from the wrong port enters the eye.
Procedure
Light Source Screen
Press the
OPM LS
soft key to display the following screen.
The wavelength and modulation mode of
the emitted light
Set the modulation mode (CW, 270Hz, 1kHz, 2kHz).
The available settings vary depending on the model. For details, see the explanation later
in this section.
Set the wavelength.
The available settings vary depending on the model. For details, see the explanation later
in this section.
Turning the Light Source On and Off
Turn the light source on after you set the wavelength and the modulation mode.
Pressing the LS Key
Press
LS
to turn on the measurement light. A mark appears on this instrument display to
indicate that the light is on.
Press
LS
while the light is on. The light turns off. The light mark disappears.
Pressing the Soft Key
Press the
LS VLS ON/OFF
soft key to display the following menu.
Turns the light on
The measurement light
turns on. A mark appears
on the AQ1100 display to
indicate that the light is on.
Turns the light off
The light turns off. The
light mark disappears.
While the above menu is displayed, you can also turn the light on and off by pressing
LS
.
Chapter 5
Light Source